
32
The following table lists the specific range parameters.
Item
Description
Valid value
Default
Calibr.
Number of days after
Disable, 1 to 7 days
Disable
Timeout
Calibration warning
(pH & ISE)
is displayed
First point
Management of 1
Replace or offset
Replace
mode (pH)
point calibration
Custom buffer
Custom buffer
Max. 5 buffers
No
(pH)
setting
pH Resolution
Set pH resolution
0.1, 0.01
0.01
display
0.001
View calibr.
Display calibration
Enable or disabled
Enable
points (pH)
points
Display out of calibr.
Enable or disable
Enable
range warning
ISE probe
Type of ISE probe
Custom or Standard (18)
Fluoride
(
HI 3222
only)
ISE unit
User, ppt, g/L, ppm, mg/L ppm
(
HI 3222
only)
ppb, M, mol/L, mmol.L, % W/V
Ion Charge
±1, ±2, none
+1
(
HI 3221
only)
To enter SETUP menu, press SETUP function key while in measuring mode.
If SETUP is not displayed press MENU key.