background image

Instruction Manual

HI142 Series

Thermologgers

Dear Customer,
Thank you for choosing a Hanna product.
Please read carefully this instruction manual before using the
logger. If you need additional technical information, do not
hesitate to e-mail us at 

[email protected] 

or

contact your nearest technical office.
These instruments are in compliance with the 

 directives.

Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.

Note:

Conserve all packing material until the instrument has
been observed to function correctly because any defec-
tive item must be returned to the Dealer in its original
packing.

PRELIMINARY EXAMINATION

GENERAL DESCRIPTION

HI142

 series is a family of temperature dataloggers especially

designed for applications in supermarket shelves and ware-
houses. It is possible to monitor continuously and simultaneously
the temperature condition in up to 31 locations.
They can all be wired through RS485 to the 

HI504903

 GSM

supervisor or to a local PC, allowing data downloading
without removing the loggers from their sites.
Moreover, with 

HI504903

 GSM supervisor it is possible to

receive SMS when an alarm occurs (low/high temperature,
low battery level, etc.) or upon user request.
The user can interact with the meters through a PC using the

HI92140

 Windows

®

 compatible software for setting High and

Low alarm thresholds, logging interval, logging delay start,
alarm mask time, lot ID, etc.
The logger can memorize up to 7600 samples (i.e. stops
when memory is full) and can also work in a cyclic way (i.e.
old data are replaced by new samples when memory is full).
In case of power failure, an internal battery assures a
working time of at least 1 week.

The status of the logger is indicated by the 2 front LEDs
(green and red) as follows:

w w w . h a n n a i n s t . c o m

WARRANTY

HI 142 loggers are warranted for one year 

against defects in

workmanship and materials when used for their intended purpose and
maintained according to instructions.
This warranty is limited to repair or replacement free of charge. Damages
due to accident, misuse, tampering or lack of prescribed maintenance
are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly packaged for
complete protection.

Hanna Instruments reserves the right to modify the design, con-
struction and appearance of its products without advance notice.

SPECIFICATIONS

Model

Range

Resolution Accuracy

HI142AH

-30.0 to 70.0°C

0.5 °C

±1.5°C

HI142BH

-10.0 to 30.0°C

0.2°C

±0.5°C

HI142CH

-30.0 to 10.0°C

0.2°C

±0.5°C

HI142DH

 20.0 to 60.0°C

0.2°C

±0.5°C

HI142EH

-30.0 to -10.0°C

0.1°C

±0.3°C

HI142FH

 20.0 to 40.0°C

0.1°C

±0.3°C

HI142GH

-5.0 to 15.0°C

0.1°C

±0.3°C

HI142HH

10 to 120°C

1°C

±2°C

Common characteristics

Probe (fixed)

1.5 m cable, food compatible

Power supply

10-20 Vdc

Dimensions

dia 86.5 x H 35 mm (3.4x1.4”)

Weight

150 g (5.54 oz.)

OPERATIONAL GUIDE

For connecting the 

HI142

 loggers with the 

HI504903

 GSM

supervisor, please refer to the

 HI504903

 instruction manual.

It is possible to connect up to 31 units within the same
RS485 network with a wires total length up to 1 km.
After connection to the GSM supervisor, the user can program
and real-time monitor the loggers and also download the
data from a PC with the 

HI92140

 software installed. Refer to

the on-line-help for further details.
The PC can be locally connected to the 

HI504903

 supervisor

with a serial cable or can call the 

HI504903

 from remote

location through a modem connection.
Before starting any operation, activate
each logger by pulling the nylon strip
in the indicated direction.
Simply hang the logger out of the
site to be monitored and insert the
temperature probe (1.5 m cable).

LED INDICATORS

TECHNICAL SERVICE

1.

Green LED always ON

: the meter

has been reset and requires new set-
ting of the parameters.

2.

Green LED flashes once every 4”

:

the meter is logging.

3.

Green LED flashes twice every 4”

:

the meter is waiting for a delayed
logging start.

4.

Green & Red LEDs flash alterna-
tively once every 4”

: the meter is

logging and an alarm condition is
occurred (at least once).

5. 

Green & Red LEDs OFF

: the meter

completed the logging procedure.

These instruments have been accurately pre-calibrated at the
factory; anyway, it is generally recommended to have all
thermometers recalibrated at least once a year.
For an accurate recalibration, contact the nearest Hanna
Customer Service Center.
If necessary, contact the nearest Hanna Customer Service
Center also for replacement of internal backup battery and/or
temperature probe.

Reviews: