Thank you for selecting a Hanna Instruments

®

 product. Please read this instruction guide before using 

the HALO probe with the Hanna Lab App or with 

HI2202

 edge

®

blu pH Bluetooth

®

 meter. For additional 

information on probe maintenance and calibration consult the Hanna Lab App “HELP” section or the edge blu 
instruction manual.

IMPORTANT! 

During servicing and measurement, handle the probe with care. Do not drop probe on a hard 

surface. It will permanently damage the probe. Utilize the clear protective cap to protect and keep the 

pH bulb hydrated during transport and storage.

PREPARATION

Remove the probe from the clear protective cap. Save the 
cap for probe storage.

1. Rinse off any storage solution or salts that may be on 

the probe’s body.

2. Verify if there is solution inside the pH bulb. Shake the 

probe down to restore continuity as the solution may 
have moved up the stem during shipping.

CONNECTION

Using Hanna Lab App

1. To begin using the HALO probe, start the Hanna Lab 

application and tap the Bluetooth icon*.

2. Press the operation button located on the top of the 

probe. The blue LED halo will begin to flash every half 
second, indicating the probe is in discoverable mode. 
The probe ID will appear in the probes list.

3. Select the probe ID from the probes list and the blue 

LED will slow down and flash every 2 seconds.

*  The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG, Inc.

Blue LED halo indicator

Operation button

Stainless steel tube