INSTRUCTION MANUAL

HI9810432

Wireless Refillable pH Tester 

for Cosmetic Creams

with built-in specialized electrode

Summary of Contents for HALO 2 HI9810432

Page 1: ...INSTRUCTION MANUAL HI9810432 Wireless Refillable pH Tester for Cosmetic Creams with built in specialized electrode ...

Page 2: ...g a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our contact list at www hannainst com Dear Customer...

Page 3: ...H Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8 10 Care Maintenance 9 Refilling the Electrode 9 Storage 9 11 Warning Error Messages 9 12 Abbreviations 9 13 Acces...

Page 4: ... 0 F Resolution pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 02 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers Hanna pH 1 68 4 01 7 01 10 01 or NIST pH 1 68 4 01 6 86 9 18 Temperature compensation Automatic ATC or Manual MTC Electrode Body material Glass Glass Low Temperature LT Junction Triple c...

Page 5: ...d data sharing Probe Features The HI9810432 features a non porous glass body that is easy to clean and withstands harsh chemicals The conical glass tip allows for penetration into semisolids and emulsions The triple ceramic junction allows a higher flow rate of electrolyte from the reference cell into the solution This high flow rate provides faster electrode response and a more stable measurement...

Page 6: ...use specified battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Navigation Press the Setup button to enter Setup mode and navigate menu items To exit Setup mode press the Setup button after SEt TIME option is displayed Press the ON OFF button to configure...

Page 7: ... in discoverable mode it will appear in the list of Available Devices Withintheapplication tap Connect topairthetesterwiththedevice Allreadingsaretransmitteddirectlytotheapplication Tag a Measurement Once connected to the application the ON OFF button can be pressed to tag the current reading Press the ON OFF button from measurement mode The display shows SEt TAG followed by TAG The reading on the...

Page 8: ...de in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 After the second point has been stored CAL SAVE is displayed briefly and the tester will return to measurement mode automatically With 3P option selected 6 Place the tip of the electrode in the second buffer rinse beaker then in the second calibra...

Page 9: ...nce junction maintains an outward flow of electrolyte Storage To ensure a quick response the glass tip and junction should be kept hydrated 1 When not in use add a few drops of storage solution to the storage cap If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Note Do not store the electrode in distilled or deionized water 2 Place the probe tip into the cap and gently ti...

Page 10: ...Electrode refilling pipette 20 pcs RECOMMENDATIONS FOR USERS Before using this product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in hazardous environments WARRANTY HI9810432 is w...

Page 11: ...ovide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interferenc...

Page 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810432 Printed in ROMANIA ...