SETUP
7. SETUP
• Short press MENU key to move to next item in menu.
• Long press MENU key to exit.
• Press arrow keys to change the values.
• Short press MENU key to automatically save modified values.
Table below presents an overview of the menu with ranges and factory set defaults.
Parameter
(Scrolled message)
Range / Option
Default settings
Control
Auto or oFF
Auto
Control type
on/oFF or Proportional
on/oFF
Control mode
Hi or Lo
Hi
Set point
4.00 to 10.00
7.20
Hysteresis (on/oFF only)
0.10 to 1.00 pH
0.50
Band (Proportional only)
0.10 to 2.00 pH
1.0
Start up delay (seconds)
0-600
60
Overtime alarm (minutes)
oFF, 0-180
30
Flow rate (Liter/hour)
Flow rate (Gallon/hour)
0.5 to 3.5
0.13 to 0.92
1.0
0.26
Level alarm
diS or En
diS
High alarm
diS or En
En
High alarm value (pH)
0 to 14.00 pH*
10.00 pH
Low alarm
diS or En
diS
Low alarm value (pH)
0 to 13.90 pH*
4.00 pH
Temperature unit
°C or °F
°C
Flow rate unit
L.H or GAL.H
L.H
Language
En (English), ES (Spanish),
Fr (French), Pt (Portuguese),
nL (Dutch), dE (German)
En (English)
*Available range changes based upon other settings.
High alarm must be set higher than low alarm. If low alarm was set 7 pH, then high alarm range is
7.1 to 14 pH.