Auto-Off

From measurement mode, press and hold the ON/OFF button. 
The LCD will show "OFF", "CAL" followed by "d08" (default 
setting, 8 minutes). Release the ON/OFF button. A single press 
on the ON/OFF button will change the auto-off timer to "d60" 
(60 minutes). To disable the auto-off feature press the ON/
OFF button again, LCD will show "d--". Press and hold to exit.

Clear Calibration

To clear the user calibration and restore the tester to factory 
default. From calibration mode, press and hold the ON/OFF 
button, the LCD will show "CLr".

“Err” Message

In calibration mode, if the probe is in the correct buffer 
solution and the "Err" message is displayed, the probe 
should be cleaned. Soak the probe in cleaning solution for 20 
minutes. Rinse with water and hydrate electrode in storage 
solution for a minimum of 30 minutes before calibrating.

Battery Indicator

When the battery level is low, the tag on the LCD will blink. 
When the battery is depleted the "Erb" message is displayed 
and the tester is powered off.

Care & Maintenance

General Information

Please read the information below, to ensure the highest possible accuracy:

• Fresh buffer should be used for each calibration, once the sachets are opened the 

buffer value can change over time.

• For improved accuracy a two-point calibration is recommended.

• If the electrode is slow or sluggish, soak it in cleaning solution for 20 minutes. 

Rinse with water and hydrate the electrode in storage solution for a minimum of 

30 minutes before calibrating.

• If measurements are taken successively, rinse the probe thoroughly in distilled or 

deionized water to eliminate cross-contamination.

• When not in use, add a few drops of storage solution to the protective cap to keep 

the glass tip and the junction hydrated. If storage solution is not available, pH 4.01 

or pH 7.01 buffer can be used. Never store the probe in distilled or deionized water.

• The outer junction sleeve can be removed and cleaned with purified water. Once 

cleaned a small amount of supplied gel electrolyte should be added to refresh the 

junction and improve the pH measurement.

Note: 

Never immerse the tester over the maximum immersion level.

Refilling the Electrode

• To remove the electrode sleeve, rotate it carefully and slide it off the probe body, keeping 

it parallel to the pH electrode.

Note: 

Handle the probe with care, the 

pH electrode stem is made of glass.

• Rinse off any traces of electrolyte gel 

from the sleeve. Soak the tip of the pH 
electrode in 

HI70061

 General purpose 

cleaning solution (or 

HI700661

HI700663

,

 HI700664

) for 20 minutes. 

Rinse with distilled or deionized water.

• Refill the reference well with 

HI9071

 

Gelled bridge electrolyte.

• Replace the sleeve making sure that the black O ring is fixed 

inside the electrode.

• Rinse off excess gel with distilled or deionized water.
• Shake the probe down as you would do with a clinical 

thermometer to eliminate any trapped air bubbles.

• Soak the probe in 

HI70300

 Electrode storage solution for a 

minimum of 30 minutes before calibrating.

Accessories

pH Buffer Solution

HI70004G

GroLine pH 4.01 buffer solution, 20 mL sachet (25 pcs.)

HI70007G

GroLine pH 7.01 buffer solution, 20 mL sachet (25 pcs.)

HI77400P

pH 4.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each)

Electrode Cleaning Solution

HI70061G

GroLine general purpose cleaning solution, 20 mL sachet (25 pcs.)

HI700661P

General purpose cleaning solution for agriculture, 20 mL sachet (25 pcs.)

HI700663P

Cleaning solution for soil deposits, 20 mL sachet (25 pcs.)

HI700664P

Cleaning solution for humus deposits, 20 mL sachet (25 pcs.)

Electrode Storage Solution

HI70300G

GroLine electrode storage solution, 20 mL sachet (25 pcs.)

HI70300-023

GroLine electrode storage solution, 230 mL

HI9072

Electrode storage solution, 13 mL dropper

Electrode Fill Solution

HI9071

Gelled bridge electrolyte

Other Accessories

HI721319

Soil auger

HI7051M

Soil preparation solution, 230 mL

Battery Replacement

To change the CR2032 lithium-ion battery, turn 

the battery cover located on the back of the tester 

counterclockwise to unlock. Remove cover and replace 

the battery with positive (+) side facing out.

Note: 

Only use the battery type specified in the 

manual. Old batteries should be disposed in 

accordance with local regulations.

Warranty

HI981030

 is warranted for a period of one year against defects in workmanship 

and materials when used for its intended purpose and maintained according to 

instructions. This warranty is limited to repair or replacement free of charge. Damage 

due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. 

If service is required, contact your local Hanna Instruments Office. If under warranty, 

report the model number, date of purchase, serial number and the nature of the 

problem. If the repair is not covered by the warranty, you will be notified of the 

charges incurred. If the instrument is to be returned to Hanna Instruments, first 

obtain a Returned Goods Authorization (RGA) number from the Technical Service 

department and then send it with shipping costs prepaid. When shipping any 

instrument, make sure it is properly packaged for complete protection.

Recommendations for Users

Before using this product, make sure it is entirely suitable for your specific application 

and for the environment in which it is used. Any variation introduced by the user to 

the supplied equipment may degrade the tester’s performance. For yours and the 

tester’s safety do not use or store the tester in hazardous environments.

Certification

All Hanna Instruments conform to the 

CE European Directives

.

Disposal of Electrical & Electronic Equipment

. The product should 

not be treated as household waste. Instead hand it over to the 

appropriate collection point for the recycling of electrical and 

electronic equipment which will conserve natural resources.

Disposal of waste batteries.

 This product contains batteries, do not 

dispose of them with other household waste. Hand them over to the 

appropriate collection point for recycling.

Ensuring proper product and battery disposal prevents potential negative 

consequences for the environment and human health. For more information, contact 

your city, your local household waste disposal service, the place of purchase or go 

to 

www.hannainst.com

.

Hanna Instruments reserves the right to modify the design, construction, or 

appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent 

of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.

IST981030 10/19

RoHS

compliant

HI981030

Soil pH Tester

INSTRUCTION MANUAL

Summary of Contents for Groline HI981030

Page 1: ...HI981030 Soil pH Tester INSTRUCTION MANUAL ...

Page 2: ...lied accessories General Description Intended Use The HI981030 Soil pH Tester is designed to measure the pH or acidity of the soil as a key factor in ensuring strong and healthy plant growth It features a single button operation system and is easy to use It has a compact and waterproof casing and automatic pH calibration at one or two points All readings are automatically compensated for temperatu...

Page 3: ...strength and resistance to ultraviolet Removable Polyvinylidene Fluoride PVDF Reference Sleeve The PVDF outer junction sleeve can be removed and cleaned Once cleaned a small amount of supplied gel electrolyte is added and the junction is refreshed improving the measurement and extending the life of the tester Specifications Range 0 00 to 12 00 pH Resolution 0 01 pH Accuracy 0 05 pH Calibration Aut...

Page 4: ...re B Note It is recommended to calibrate the electrode with buffers at the temperature it will be used at A B A One or Two Point Calibration with pH 7 01 When 7 01 is displayed place the tip of the electrode in pH 7 01 buffer When the reading is stable the stability icon will disappear When 4 01 is displayed follow the procedure to the right for a one or two point calibration One Point Press the O...

Page 5: ...he Erb message is displayed and the tester is powered off Care Maintenance General Information Please read the information below to ensure the highest possible accuracy Fresh buffer should be used for each calibration once the sachets are opened the buffer value can change over time For improved accuracy a two point calibration is recommended If the electrode is slow or sluggish soak it in cleanin...

Page 6: ... trapped air bubbles Soak the probe in HI70300 Electrode storage solution for a minimum of 30 minutes before calibrating Accessories pH Buffer Solution HI70004G GroLine pH 4 01 buffer solution 20 mL sachet 25 pcs HI70007G GroLine pH 7 01 buffer solution 20 mL sachet 25 pcs HI77400P pH 4 01 7 01 buffer solution 20 mL sachet 10 pcs 5 each Electrode Cleaning Solution HI70061G GroLine general purpose ...

Page 7: ...mendations for Users Before using this product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For yours and the tester s safety do not use or store the tester in hazardous environments Certification All Hanna Instruments conform to the CE ...

Reviews: