Hanna Instruments reserves the right to modify the design, con-
struction and appearance of its products without advance notice.

WARRANTY

HI 991404

 and 

HI 991405 

are 

warranted for two years 

against defects in

workmanship and materials when used for their intended
purpose and maintained according to instructions. 

The probes

are warranted for a period of six months

. This warranty is limited to

repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of
prescribed maintenance are not covered.
If service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the
model number, date of purchase, serial number and the
nature of the failure. If the repair is not covered by the
warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization Number from
the Customer Service department and then send it with
shipment costs prepaid. When shipping any instrument,
make sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is
prohibited without the written consent of the copyright
owner, Hanna Instruments Inc.

Instruction Manual

(HI 991404 - HI 991405)

Dear Customer,
Thank you for choosing a Hanna product. This
manual will provide you with the necessary in-
formation for a correct operation. Please read it
carefully before using the meter. If you need
additional technical information, do not hesitate
to e-mail us at [email protected]

[email protected]

[email protected]

[email protected]

[email protected].

These instruments are in compliance with the

directives.

Remove the instrument from the packing mate-

rial and examine it carefully. If any damage has

occurred during shipment, immediately notify

your Dealer or the nearest Hanna Customer

Service Center.
The meters are supplied with:

 pH 4.01 and 7.01 buffer solutions (20 mL each);

• 

HI 70031 

1413 µS/cm calibration solution, for 

HI

991404

;

• 

HI 70030 

12.88 mS/cm calibration solution, for 

HI

991405

;

• 

HI 1293D

 pH-electrode with differential input signal;

• Hanna 12 VDC power adapter;
• Instruction manual.

Note: Conserve all packing material until the

instrument has been observed to function

correctly. Any defective item must be re-

turned in its original packing.

PRELIMINARY EXAMINATION

COMBO 

 

is a combined pH/tempera-

ture/EC/TDS meter specially designed to

meet the needs of growers in greenhouses

and hydroponic applications.
You can simply install the meter above the

sample to be tested for continuous

measurement.
The 

HI 1293D

 pH-electrode & the EC/TDS/

Temperature probe have been molded with

pipe thread (1/2’’ NPT), which allows the

user to attach them to an in-line system.
Measurements are accurate and the meter

can be calibrated at one or two points for

pH and at single point for EC/TDS.
You no longer need to worry about battery

changes either: the unit runs without

interruption on 12 VDC power supply.

GENERAL DESCRIPTION

FUNCTIONAL DESCRIPTION

These Instruments are in

Compliance with CE

Directives

SPECIFICATIONS

Recommendations for Users

Before using this product, make sure that it is entirely

suitable for the environment in which it is used. Operation

of this instrument in residential areas could cause unac-

ceptable interferences to radio and TV equipment.

The glass bulb at the end of the electrode is sensitive to

electrostatic discharges. Avoid touching this glass bulb at all

times. During operation, ESD wrist straps should be worn to

avoid possible damage to the electrode by electrostatic

discharges.

Any variation introduced by the user to the supplied equip-

ment may degrade the instrument’s EMC performance.

To avoid electrical shock, do not use this instrument when

voltages at the measurement surface exceed 24 VAC or 60

VDC. To avoid damages or burns, do not perform any

measurement in microwave ovens.

ISTR991404R1    05/02

HI 991404

EC/TDS Range

0 to 3999 µS/cm

0 to 2000 ppm

Resolution

1 µS/cm - 1 ppm

Accuracy (@20ºC/68ºF)

±2% f.s.

Typical EMC Deviation

±2.5 f.s.

HI 991405

EC/TDS Range

0.00 to 20.00 mS/cm

0.00 to 10.00 ppt

Resolution

0.01 mS/cm - 0.01 ppt

Accuracy (@20ºC/68ºF)

±2% f.s.

Typical EMC deviation

±2.5% f.s.

Common specifications

pH Range

0.0 to 14.0 pH

Temperature Range

0.0 to 60ºC (32.0 to 140ºF)

Resolution

0.1 pH

0.1ºC or 0.1ºF

Accuracy (@20ºC/68ºF)

±0.1 pH

±0.5 ºC or ±1ºF

Typical EMC Deviation

±0.1 pH

±1 ºC or ±2 ºF

Temp.Compensation     pH

Automatic

                                EC/TDS

β

=0.0 to 2.4%/ºC

TDS Conversion factor

0.45 to 1.00 (CONV)

pH electrode

HI 1293D

 (

included

)

pH Calibration

At 1 or 2 points

with auto-buffer recognition

EC/TDS Calibration

Automatic at 1 point

Power supply

   12VDC power adapter (

included

)

Environment

0 to 50ºC (32 to 122ºF);
RH 95% non-condensing

Dimensions (meter only)

160 x 105 x 31 mm

(6.2 x 4.1 x 1.2’’)

Weight (meter only)

190 g (6.7 oz.)

1. Liquid Crystal Display
2. ON/OFF/MODE button
3. Power supply connector
4. 12 VDC power adapter
5.

HI 1293D 

pH-electrode with differential

input, pipe thread 1/2’’ NPT

6. EC/TDS/Temperature probe, pipe

thread 1/2’’ NPT
(works also as matching pin for pH-

electrode)

7. DIN connector
8. SET/HOLD button