Instruction Manual

HI 96708

Nitrite High Range ISM

w w w . h a n n a i n s t . c o m

[email protected]

www.ponpe.com

Summary of Contents for 96708

Page 1: ...1 Instruction Manual HI 96708 Nitrite High Range ISM w w w h a n n a i n s t c o m sale ponpe com www ponpe com...

Page 2: ...1 BATTERY REPLACEMENT 22 ACCESSORIES 22 WARRANTY 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrume...

Page 3: ...Selective Meter is supplied complete with Two Sample Cuvettes and Caps CAL CHECK standard cuvettes 9V Battery Scissors Cloth for wiping cuvettes Instrument quality certificate Instruction Manual Rigi...

Page 4: ...irt by a transparent cup This makes the instruments fulfill field applications Display messages aid the user in routine operation The meter has an auto shut off feature that will turn off the instrume...

Page 5: ...last reading reminder Dimensions 192 x 102 x 67 mm 7 6 x 4 x 2 6 Weight 290 g 10 oz REQUIRED REAGENTS Code Description Quantity test HI 93708 0 Powder Reagent 1 packet Precision Precision Precision Pr...

Page 6: ...c molar concentration of the substance d optical path through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the other factors are known Photomet...

Page 7: ...nto the desired measuring unit and to display it on the LCD The measurement process is carried out in two phases first the meter is zeroed and then the actual measurement is performed The cuvette has...

Page 8: ...ODES Measurement mode default operation mode enables both validation and measurement Calibration mode may be entered by keeping CAL CHECK pressed for three seconds the CAL tag appears it enables calib...

Page 9: ...lease check the preparation of the zero cuvette Light Low There is not enough light to perform a measurement Please check the preparation of the zero cuvette No Light The instrument cannot adjust the...

Page 10: ...dition The concentration of the sample is beyond the programmed range dilute the sample and re run the test c during calibration procedure Standard Low The standard reading is less than expected Stand...

Page 11: ...k Proper use of the powder reagent packet a use scissors to open the powder packet b push the edges of the packet to form a spout c pour out the content of the packet It is important that the sample d...

Page 12: ...imes reported in this manual are referred to 20 C 68 F As a general rule of thumb they should be doubled at 10 C 50 F and halved at 30 C 86 F In order to maximize accuracy prior to a measurement follo...

Page 13: ...ple up to the mark and replace the cap Place the cuvette into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press ZERO CFM and the lamp cuvette and det...

Page 14: ...n period Alternatively wait for 10 minutes then just press READ TIMER In both cases the lamp cuvette and detector icons will appear on the display depending on the measurement phase At the end of meas...

Page 15: ...ssing ON OFF When the beeper sounds briefly and the LCD displays dashes the meter is ready Place the CAL CHECKTM Standard HI 96708 11 Cuvette A into the cuvette holder and ensure that the notch on the...

Page 16: ...he display depending on the measurement phase At the end of the measurement the display will show the validation standard value The reading should be within specifications as reported in the CAL CHECK...

Page 17: ...r sounds briefly and the LCD displays dashes the meter is ready Press and hold CAL CHECK for three seconds to enter calibration mode The display will show CAL during calibration procedure The blinking...

Page 18: ...ing on the measurement phase After measurement the instrument will show for three seconds the CAL CHECKTM Standard value Note If the display shows STD HIGH the standard value was too high If the displ...

Page 19: ...umber is automatically increased When the correct month has been set press ZERO CFM or READ TIMER to confirm Now the display will show the day blinking Press GLP to edit the desired day 01 31 If the k...

Page 20: ...isplay and the instrument returns to measurement mode after three seconds It is possible to delete the calibration and restore factory calibration Press GLP to enter GLP mode In the GLP mode the last...

Page 21: ...yed before auto shut off the value is displayed when the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 750 measurements depen...

Page 22: ...steps ACCESORIES REAGENT SET REAGENT SET REAGENT SET REAGENT SET REAGENT SET HI93708 01 Reagentsfor100tests HI93708 03 Reagentsfor300tests OTHER ACCESORIES OTHER ACCESORIES OTHER ACCESORIES OTHER ACCE...

Page 23: ...ng to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If service is require...