Quick Guide
Supra R30
Cat. No. : SU-R30
Display & Control Panel
Setting Temp. Limit
Start
Stop
Key Lock Power Switch
Lid Open
Max. RPM :
30,000 rpm
Max. RCF :
106,657 xg
Setting Temperature
Setting temperature ( -10℃ ~ 40℃)
Setting temp. limit
Setting Time
Up to < 100 hr, Continuous run (00:00)
LED Display
Status of RPM / RCF / Time /
Temperature / Rotor ID /
Program / ACC / DEC
Setting Rotor ID.
View Settings
Verifying the setting
parameters during
centrifugation
Pulse
After reaching the set speed,
the centrifugation is decelated
and stopped
Setting Speed
Setting RPM
Setting RCF
Setting ACC / DEC Rate
ACC (0 ~ 9 steps, Max.9)
DEC (0 ~ 9 steps, Natural DEC.0)
Program Save / Call
Saving a program
(up to 100)
Calling up a program
Saving / Calling Up a Program
Press a parameter
to be changed
Press the
Enter button
Press the
Enter button
Enter the desired valuse
on the keypad
Enter the desired program
number on the keypad
Press the
Save or call button
Setting Run Parameters
(RPM / RCF / Time / Temperature / ACC / DEC)
Press the
Enter button
Press the
Temp. limit button
Press the
Enter button
Temperature range that stops
operation when the temperature
is lower or higher than the set
temperature.
Enter the desired temperature
range on the keypad
After loading the rotor,
press the Rotor button
Enter the rotor ID
on the keypad
Setting Rotor ID
Serial No.
Installation Date