TROUBLESHOOTING 

 

The following table lists Megapulse

®

 

problem symptoms and possible areas to check for the problem causes. If these 

suggested measures do not correct the machine malfunction, call your ACP Customer Support representative. 
 

PROBLEM 

CAUSE 

REMEDY 

Unit will not power on 

  Power Cord not plugged into the 

unit or AC outlet 

  Verify the electrical cord has a molded 

HOSPITAL 

GRADE

 plug (see below) 

  Verify if power cord is connected as appropriate and 

check that the cord is not frayed/damaged 

  Verify AC outlet is functional 

APPL message is 
displayed on the unit 

  The drum inductive applicator is 

not connected to the generator 
unit. 

  Verify that the drum inductive applicator connector is 

plugged into the back of the main unit.  Check that the 
connectors are plugged-in properly. 

  If the connector is plugged in properly, and upon turning 

off the unit and re-starting, the same message is 
displayed, please contact the Service Center. 

FAIL message is 
displayed on the unit 

  There is an internal failure in the 

unit 

  Contact the Service Center 

HOT message is 
displayed on the unit. 

  The output power may be out of 

appropriate calibration range 

  Contact the Service Center 

Treatment will not start 

  No program selected 

  Select desired treatment program and press START 

Patient cannot detect 
output, when unit is ON 
and treatment running 

  drum inductive applicator is not 

emitting output 

  Patient has impaired sensation 

over treatment area 

Use output tester (if available) to determine if unit has failed 

or is operating incorrectly.  

  If LED does not illuminate properly, contact ACP at (800) 

350-1100 about replacing the unit 

  If LED is illuminated, discontinue treatment and test 

patient for impaired sensation over treatment area.  

 
 
 
 
 
 
 
 

 

Calibration 

 

If  clinicians  are  to  give  prescribed  therapeutic  doses  of  shortwave  with  reasonable  accuracy,  the  machine  used  is 
required  to  meet  the  equipment  performance  standard  (21  CFR  890.5290)  instituted  by  the  Food  and  Drug 
Administration.  Periodic calibration of treatment parameters having a direct influence on the physiological effect of 
shortwave  on  tissue,  i.e.,  power  output,  and  time  accuracy,  is  a  major  part  of  this  regulation.    It  is  suggested, 
therefore, that the Megapulse

®

 

be verified at least once yearly.  It is recommended that recalibration be performed by 

qualified service personnel. 

 

Due  to  the  sophisticated  nature  of  the  Megapulse

®

,  ACP  requires  that  all  service  work  be  performed  at  its 

manufacturing or service facility or by a qualified ACP field technician or ACP authorized service facility. 

 

Service Center 

 

For repair or service of ACP products and accessories, please call (800) 350-1100 and follow the prompts. Normal 
hours  of  operation  are  6:00am  to  5:00pm  Pacific  Standard  Time.  After  hours,  please  leave  a  message  and  a 
technician will return your call during the next scheduled workday. 

Identify  the  “HOSPITAL 
GRADE” marking on the 
power plug

 

 

Summary of Contents for ACP MEGAPULSE II

Page 1: ...by reference in its entirety This Manual may not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual wit...

Page 2: ...Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNICYCLE OMNIVR and SYNCHRONY represent the most recent worldwide advances available for therapeutic application of electromedical devices and othe...

Page 3: ...ts Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust accidental c...

Page 4: ...olation 12 Cleaning Disinfecting of the Megapulse II 12 Cleaning and Low Level Disinfection 12 Intermediate Level Disinfection 13 Use of Barriers 13 MODES OF OPERATION 14 Thermal Effect Variables 14 P...

Page 5: ...healthcare professional Indications Diathermy is used therapeutically to increase the temperature of tissues It is generally accepted that heat produces the following desirable therapeutic effects Rel...

Page 6: ...ad or any implanted system that may or may not contain a lead Both the thermal and sub thermal modes of operation pose a risk of tissue destruction o If you are a physician who implants or monitors pa...

Page 7: ...y over the cranial region as little is known about the electromagnetic effects on the cerebrum to determine if it represents a serious hazard when shortwave diathermy is applied to the head Do not app...

Page 8: ...le and a mattress or couch without metal parts e g decorative buttons springs near the patient applicators or cables Do not use conductive mattresses or mattress covers During application of shortwave...

Page 9: ...time and if there is evidence of or if the patient complains of pain or overheating during treatment adjust the output downward until it is tolerated by the patient If the patient continues to complai...

Page 10: ...ether understood it is well established that this energy causes vibration of ions and molecules within the cells and tissues Depending on the intensity of the high frequency energy applied this vibrat...

Page 11: ...indicated by the output display LED G Pulse duration switches select the pulse duration in microseconds as displayed in the PULSE DURATION display Pressing the Up Arrow increases the pulse duration to...

Page 12: ...commonly used pulsed shortwave electrode Drum inductive applicators are generally referred to as monodes Drum inductive applicators are operated in the pulsed mode to avoid overheating of the coils Hi...

Page 13: ...a depth of 2 inches 5cm Thermal Dosage Pulsed shortwave diathermy provides outputs capable of producing thermal effects Average power output in watts is controlled by the peak power and the duty facto...

Page 14: ...nfecting of the Megapulse II Modality equipment shall be cleaned disinfected per facility infection control policy ACP recommends the following guidelines Proper cleaning procedures at the end of each...

Page 15: ...ed to cover any surface that may be touched during a patient treatment to help prevent cross contamination Barrier film is for single use only The film is discarded after each patient treatment The pr...

Page 16: ...e field applicators Typical values for the Megapulse II drum inductive applicator is 1 5 SAR The higher the SAR value the greater the thermal effect in the tissue III Treatment Time In general treatme...

Page 17: ...ation following stretch or mobilization making PSWD an effective modality for treatment of scar tissue and contracture Summary of Physiologic Effects 1 Vasodilatation of blood vessels 2 Increase in ce...

Page 18: ...show 800 PPS If the OUTPUT ON OFF CONTROL 6 switch is pressed with no time set on the display the buzzer will sound and the output will not be energized 3 Audible confirmation of the operation of the...

Page 19: ...t contact 5 When the drum inductive applicator has been placed in position tighten the hand wheels on the arm to prevent movement Switch treatment ON 6 The protocol LED 3 on the front panel will light...

Page 20: ...press gently on the drum applicator pin to dislodge the drum applicator from the articulating arm The drum applicator can now be applied to a patient where the articulating arm might have obstructed t...

Page 21: ...en unit is ON and treatment running drum inductive applicator is not emitting output Patient has impaired sensation over treatment area Use output tester if available to determine if unit has failed o...

Page 22: ...to May of 2004 the default pulse rate setting is 100 PPS Pulse Duration 400 200 100 65 40 20 s available Pulse Repetition Rate 50 100 200 400 600 800 PPS available Timer The digital treatment timer in...

Page 23: ...ic Representation of Pulse Trains 20 MICROSECOND Sec PULSES 800 PULSES PER SECOND PPS 65 Sec PULSES 400 PPS 200 Sec PULSES 200 PPS 400 Sec PULSES 100 PPS CONTINUOUS TRAIN OF PULSES PULSES CONTAIN 27 1...

Page 24: ...R IN WATTS W EFFECT 400 Sec 3 0W 6 0W 12 0W 24 0W 36 0W 48 0W 4o C 200 Sec 1 5W 3 0W 6 0W 12 0W 18 0W 24 0W 2o C 100 Sec 75W 1 5W 3 0W 6 0W 9 0W 12 0W 1o C 65 Sec 49W 1 0W 2 0W 3 9W 5 9W 7 8W SUBTHERM...

Page 25: ...29000 Megapulse Power Cord Hospital Grade 10ft 72763 Applicator Head Coaxial Cable 89555 Megapulse II User Manual Infection Control Supplies ITEM ITEM NO DESCRIPTION 55536 Super Sani Cloth Wipes Sing...

Page 26: ...f any of the following have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been tamper...

Page 27: ...nty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipped complete with all accessories exce...

Page 28: ......

Reviews: