Operating Instructions
Numbers in brackets in the following operating instructions refer to the front panel drawing FIGURE below.
1.
Connect the Megapulse
®
II
to a suitable AC outlet, and energize the unit by pressing the Power Switch (1) up.
The Power Switch will light up, the buzzer will sound briefly and the Timer Display (2) will show -65-. The
microprocessor in the unit now completes a short self-test routine.
2.
After 4 seconds the Timer Display (2) will change to 20:00. The 1
T MODE (3) indicator LED will light, the
PULSE DURATION DISPLAY (4) will indicate 100
s and the PULSE RATE DISPLAY (5) will show 800
PPS. If the OUTPUT ON/OFF CONTROL (6) switch is pressed with no time set on the display, the buzzer will
sound and the output will not be energized.
3.
Audible confirmation of the operation of the front panel push switches may be enabled by turning the unit on
while holding down the VAR Mode switch (3). The display will read “set up” then “beep on” or “beep off”.
Use the PULSE DURATION
(7) key to turn the beep to “on” or “off”. Press a timer key (8) to exit setup and
save the settings.
4.
Set the required treatment time by pressing the timer switches (8). The maximum treatment time allowed is 30
minutes. This can only be adjusted if the variable (VAR) MODE has been selected.
5.
The connecting cables associated with the drum inductive applicators should be positioned in such a way that
contact with a patient, or conductive or energy absorbing object, is avoided.