Troubleshooting
p. 18
User gUiDe
ceiling motor
9. Troubleshooting
9.1. General
the ceiling motor is not working at all:
■
Check if by any chance the emergency switch is activated and the motor is turned off (on/off LED is not lit).
Deactivate the emergency switch (on/off LED illuminates).
■
Check if the batteries are charged.
■
Try the buttons on the Control Panel. If they are working, there is a problem with the handcontrol or the cable.
• 2805 - 2815 L IR: the 3 1,5 V AAA batteries in the handcontrol could be empty.
• 2800 - 2805 L: Check if the spiral cable is correctly attached to the connector.
When using the handcontrol, you hear a beeping sound and/or the battery leD on the control panel lits
blue:
■
The batteries have to be charged urgently.
■
If the ceiling motor does not work, call the technical service.
9.2. Removing the cover
■
Hold the white side cover near the logo and softly pull the cover.
■
Place a tool (that fits the hole) in the opening and carefully press
the internal clip until the cover becomes undone. Repeat this for the
other side of the white cover. Remove the cover.
■
If required, repeat this for the second white cover.
■
With regard to the cover with the contact strips, make sure to dis-
connect the connector from the board.
9.3. Assembling the cover
■
Position the white cover facing the gray cover
■
With regard to the cover with the contact strips, make sure to recon-
nect the connector to the board.