Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
[SENSe:]FREQuency:APERture? [{MIN|MAX}]
Returns the gate time of the frequency measurement function in AC V / AC I mode.
Return values:
<GateTime>
1.00000000E-02, 1.00000000E-01, 1.0000
MIN: 1.00000000E-02
MAX: 1.0000
[SENSe:]FREQuency:CURRent:RANGe:AUTO
<
Mode
>
Activates or disables the auto range mode of the AC I measurement function for frequency measurements.
Parameters:
<Mode>
ON | OFF | 1 | 0
ON:
Activates the auto range mode of the AC I measurement function
OFF:
Disables the auto range mode of the AC I measurement function
1:
Activates the auto range mode of the AC I measurement function
0:
Disables the auto range mode of the AC I measurement function
*RST: ON
[SENSe:]FREQuency:CURRent:RANGe:AUTO?
Returns the auto range state of the AC I measurement function for frequency measurements.
Parameters:
<Mode>
1 | 0
1:
Auto range mode of the AC I measurement function is activated
0:
Auto range mode of the AC I measurement function is disabled
[SENSe:]FREQuency:CURRent:RANGe[:UPPer] {
<
Current_Range
>
|MIN|MAX|DEF}
Sets the current range of the AC I measurement function for frequency measurements.
Parameters:
<Current_Range>
20mA, 200mA, 2A, 10A
MIN: 20mA
MAX: 10A
DEFault: 20mA
*RST: AUTO
[SENSe:]FREQuency:CURRent:RANGe[:UPPer]? [{MIN|MAX}]
Returns the current range of the AC I measurement function for frequency measurements.
Return values:
<Current_Range>
20mA, 200mA, 2A, 10A
MIN: 2.00000000E-02
MAX: 1.0001
DEFault: 2.00000000E-02