background image

 

 

19

 

 

Setting up the Equipment  

 

 

This method allows centering to .0005 in. (0.0127 mm). 

 

1.

 

Adjust the A-514 Self-Centering Bore adapters for both 

the laser and target. 

Using the L-514GS or A-514GL Leg-Setting Gage, set the 

adapter legs to the desired bore radius. See 

Setting the Legs 

for the A-514GS

 on Page 12 and 

Setting the Legs for the A-

514GL

 on Page 12 for more information on this procedure. 

2.

 

Insert the L-708 Laser and A-512 Target into their 

adapters. 

When using the small adapter, hold the laser from the back 

end and tip it slightly upward, inserting the adapter and 

laser at an angle. The weight of the laser will pull it down 

after releasing. The medium and large bore adapters are 

provided with counterweights to hold the laser in place so 

the unit doesn’t tip.    

3.

 

Insert the laser and target assemblies into the two 

reference bores. 

Attach the weighted insertion rod to the target/adapter 

and tip the unit forward to insert into the bore. The 

weight of the handle forces the target to tip backward and 

seat the  target legs into the bore, centering the 

adapter/target.  Ensure that the target cable is at the 6 

o’clock position and release the handle as soon as the 

target is in place in the bore end. The target centers itself 

securely.  

Repeat for the laser/adapter assembly. 

Note:

 

To remove the target, lift the handle up and slide the target/adapter out of the bore.  

4.

 

Use the laser’s angular adjustments to set/tilt the laser to zero on the target, establishing the 

reference bore centerline. 

 

 

 

 

Figure 13 – 

Laser mounted in bore adapter 

Figure 14 – 

Inserting the Target 

Summary of Contents for L-708

Page 1: ......

Page 2: ...ticles of equipment manufactured by other companies and limited by a warranty extending for less than one year In such an event the more limited warranty applies to said instrument or article of equip...

Page 3: ...S and A 514GL 11 Setting the Legs for the A 514GS Using Small and Medium Bore Adapters 12 Setting the Legs for the A 514GL Using Large Bore Adapters 13 The Model R 1307 Readouts 14 Model R 1307C 14 Mo...

Page 4: ...ing error of up to 002 in 0 05 mm can occur In addition the laser switch setting CONT or Fixed vs PULSE must also agree with the R 1307 Readout setting For CONTinuous Fixed Mode set the readout to F10...

Page 5: ...to the desired bore radius using the Leg Setting Gage The L 708 Laser and A 512 Target are inserted into their Bore Adapters The entire laser and target assemblies are inserted into the two reference...

Page 6: ...der good conditions should hold 010 0 25 mm over the full distance The L 708LD requires the use of the T 1218 2 Axis Target T 1220 or T 1240 targets 20x20 mm PSDs since the laser beam diameter is larg...

Page 7: ...Adjustments is a small window containing a white dot The white dot follows the coarse angular adjustment of both axes If the white dot displays in the center of the window both the Horizontal and Vert...

Page 8: ...t establishing a reference bore centerline Pulse Continuous Modes for the L 708 Laser The L 708 Laser is equipped with a PULSE CONTinuous switch which manually switches the laser beam between Pulsed a...

Page 9: ...and remove the cover 3 Replace the two batteries Remove the old batteries and replace them with new 9 volt cells being careful to orient them with the negative terminal out up 4 Re attach the cover R...

Page 10: ...he counter weight to the poles 4 An extra counter weight and pole are included with the A 514C Insert the pole in the main counterweight and use the thumbscrew to tighten 5 Slide the extra counterweig...

Page 11: ...t and adapter in self centering mode are inserted into the reference bore A measurement is taken with the target in both the NORMal and INverted positions see Appendix A The NORMIN Method beginning on...

Page 12: ...for bores from 6 5 in 165 mm to 17 5 in 445 mm A 514C for bores from 17 0 in 432 mm to 40 0 in 1 m Note For bores over 1 m half bores or where bore surfaces are worn or rough the T 218 Two Axis Univer...

Page 13: ...e Adapters The A 514A B and C Bore Adapters are all assembled the same way Each adapter consists of two legs that cross each other forming an X shape Using a bolt and Allen wrench provided screw the b...

Page 14: ...S is used with the small A 514A or medium A 514B bore adapters The A 514GL can be used for all three adapters and must be purchased if using the large bore adapter Figure 6 The A 514GS left and A 514G...

Page 15: ...e the stop pin and turn the bore adapter one more time replacing the stop pin Repeat the process for the next leg 9 To set the remaining two legs the bore adapter must be removed from the hub and flip...

Page 16: ...e other end in the slot 3 Set the caliper to 7 in and press one end of the caliper against the pin of the extension 4 Slide the leg setting plate up against the other end of the caliper 5 After the le...

Page 17: ...07C Supports cabled local targets only Supports both pulsed beam and continuous laser modes Functional replacement for the R 307 Analog Readout Model R 1307 900 2 4 and R 1307 2 4ZB Supports both wire...

Page 18: ...r beam to target position information and plot results that can be saved and printed The R 358 computer interface provides exceptionally high accuracy 00004 in resolution for downloading target data t...

Page 19: ...16 Model R 1307 Readout Control Panel...

Page 20: ...17 Model R 1307C Readout Control Panel...

Page 21: ...a contraction of NORMal INverted which briefly describes the method In Step 3 Laser Setup on screen instructions guide the user through setting up the laser and making it parallel to reference points...

Page 22: ...re provided with counterweights to hold the laser in place so the unit doesn t tip 3 Insert the laser and target assemblies into the two reference bores Attach the weighted insertion rod to the target...

Page 23: ...il the upper display shows Use the UP and DOWN arrow keys to select 5 Select the PSD descriptor applicable to your target Press the MENU button until the upper display shows Use the UP and DOWN arrow...

Page 24: ...E READOUT AS FOLLOWS R 1307C SET FUNCT TO LOCAL ID AND CH NOT USED R 1307 SET FUNCT TO LOCAL SET ID OF UPPER AND LOWER DISPLAY TO THE SAME VALUE DEFAULT 1 SET CH TO A VALUE NOT USED BY ANOTHER READOUT...

Page 25: ...ontal X axis position of the local target 4 Set the System ID Radio Channel Press the MENU button until the upper display shows with the current System ID blinking Use the UP and DOWN arrow keys to se...

Page 26: ...ed in bore alignments the True Bore Center the Target Center and the Laser Reference Centerline If mounting fixtures were perfect the Target Center would be located at the True Bore Center and if perf...

Page 27: ...on the readout are also rotated Therefore although the Target Center appears to be to the right of and lower than the Laser Beam Center in Figure 18 the vertical readings are positive and the horizont...

Page 28: ...sured point is the Measured Diameter TCE MDT The formula is MDT Normal Reading Inverted Reading 2 3 Calculate the difference in diameter between each measured point and the reference location The diff...

Page 29: ...are provided as well as a sample plot of readout data see Figure 20 on Page 27 LOCATION NORMAL INVERTED OFFSET DIAMETER Reference Location Measuring Point 1 Measuring Point 2 Measuring Point 3 Measur...

Page 30: ...27 Figure 20 Sample plot of readout data...

Page 31: ...01 in ft 0008 mm m Beam Stability 0001 in hr F 0 004 mm hr C 0001 in ft hr F 0 01 mm M hr C Laser Type Diode Laser 0 9 mW Cw BRH Class II Laser Wavelength 670 nanometers Materials Aluminum Flange All...

Page 32: ...29...

Page 33: ...the type intended for nail polish removal NOTE When cleaning precision optics even with the best quality optical cleaning tissue use gentle pressure to avoid scratching the surface or damaging the opt...