background image

33 

 

Appendix D – Care and Cleaning of Target Optics 

 

The proper care and cleaning of optical windows and/or lenses of Hamar Laser’s position-sensing devices 

(targets) assures optimum performance. Contaminants on an optical surface increase scatter, absorb laser 

energy, and eventually degrade the accuracy of the position-sensing devices. Because cleaning any 

precision optic risks damaging the surface, optics should only be cleaned when absolutely necessary. 

When cleaning is required, we recommend the following supplies and procedures. 

 

Required Supplies 

 

 

Optics Cleaning Tissue:

 Soft, absorbent, lint-free lens tissue 

 

Swabs:

 Cotton swabs with wooden handles or polyester swabs with polypropylene handles 

 

Dust Blower:

 Filtered dry nitrogen blown through an antistatic nozzle is best. Canned dusters, such 

as Dust-Off, will also work.  

 

Mild Soap solution:

 Neutral soap, 1 percent in distilled water. Avoid scented, alkali, or colored soap 

such as liquid dishwashing detergents or hand soap. Ten drops of green soap (available at a 

pharmacies and optical cleaning suppliers) per 100 cc of distilled water is an acceptable alternative. 

 

Isopropyl Alcohol:

 Spectroscopic grade. Over-the-counter alcohol contains too much water and may 

have impurities. 

 

Acetone:

 Spectroscopic grade. Do not use over-the-counter Acetone, such as the type intended for 

nail polish removal. 

 

NOTE:

 

When cleaning precision optics, even with the best  quality  optical  cleaning tissue, use gentle 

pressure to avoid scratching the surface or damaging the optical coating(s). Always wipe using a figure-

eight motion in one direction (begin at the top and work toward the bottom in a figure-eight motion). 

Use only moistened (not soaked) optical cleaning tissue, Swabs and Spectroscopic grade Acetone and 

Isopropyl Alcohol. Never spray any type of liquid directly on the device or submerge any part of the 

device. 

 

Removing Dust 

 

Dust can bind to optics by static electricity. Blowing only removes some of the dirt. The remainder can be 

collected by using wet alcohol and Acetone swabs wrapped with optical lens tissue. Acetone dries rapidly 

and helps to eliminate streaks.  

 

1.

 

Blow off dust. 

2.

 

If any dust remains, twist lens tissue around a cotton swab moistened in alcohol and repeat as 

necessary. 

3.

 

Repeat using Acetone. 

 

Cleaning Heavy Contamination 

 

Fingerprints, oil, or water spots should be cleaned immediately. Skin acids attack coatings and glass and 

can leave permanent stains. Cleaning with solvents alone tends to redistribute grime.   

 

1.

 

Blow off dust.  

2.

 

Using a soap-saturated lens tissue around a swab, wipe the optic gently. Repeat as necessary. 

3.

 

Repeat using a distilled water-saturated lens tissue wrapped around a swab. 

4.

 

Repeat using an alcohol-saturated lens tissue wrapped around a swab.  

5.

 

Repeat using an acetone-saturated lens tissue wrapped around a swab. 

 

Summary of Contents for Dual Scan L-732

Page 1: ......

Page 2: ...ing the alignment measurements For best results do not start taking measurements until the recommended warm time has passed It is always a good metrology practice to periodically check your reference...

Page 3: ...ticles of equipment manufactured by other companies and limited by a warranty extending for less than one year In such an event the more limited warranty applies to said instrument or article of equip...

Page 4: ...al Pitch Axis 13 Checking the Levels for Accuracy 13 Measurement Overview 14 Selecting Reference Points on Machine Tools 14 Selecting Reference Points on Process Mills 14 Measurement Summary 15 Measur...

Page 5: ...sfer line wing bases Injection molding machines and presses Aircraft assembly seat track alignment setting water butt and station planes wing to body and body to body join alignment etc Roll paralleli...

Page 6: ...and targets fit into a small portable shipping case Diode laser 2 times more stable than HeNe based laser systems Battery or AC powered Laser planes flat to arc seconds 00003 in ft or 0 0025mm m in 18...

Page 7: ...to body and body to body join alignment etc Saw mills Vertical press alignment Measuring and aligning The flatness of almost any horizontal or vertical surface squares frames ways flanges circles etc...

Page 8: ...p times Applications for alignments with tolerances of 00002 in ft or 0 0017 mm m or higher Roll parallelism in paper mills printing presses and film lines Leveling almost any surface squares frames w...

Page 9: ...m stability Applications for alignments with tolerances of 00015 in ft or 0 01 mm m or greater Normal or blown film lines roll alignment Rubber roll alignment Leveling almost any surface squares frame...

Page 10: ...ine adjustments see Figure 1 Number of laser planes the L 743 and the L 733 have three laser planes and the L 742 and L 732 have two laser planes that can be factory configured for two vertical planes...

Page 11: ...parallel to your references However doing this may result in some laser drift Accordingly if you use the laser prior to the warm up period it is recommended that you go back and check your references...

Page 12: ...on speed control This control slows the scanner spin until you can see the laser beam pass over the target the farther away the target is located the slower the turret must spin Figure 3 shows control...

Page 13: ...ure calibrates only one level vial at a time and must be repeated for the other axis Note 2 It is very important to warm up the laser for at least 30 minutes before starting this procedure It is also...

Page 14: ...alibration of the Level Vials Roll Axis Note It is very important to warm up the laser for at least 30 minutes before starting this procedure It is also very important to level both the Pitch and Roll...

Page 15: ...tance between the points The result is the calibration of the level vial in units of in ft or mm m To be within specifications the calibration result should be as follows Split Prism Vial Assembly 000...

Page 16: ...ulating the Calibration of the Levels Vials Pitch Axis 1 Level the laser Rotate the entire laser unit 90 to calibrate the Pitch Axis level vial Using the adjustment knobs level the laser so that both...

Page 17: ...Pitch Axis level you are adjusting Using the wrench provided adjust the level assembly until the bubble is centered in the window for the Standard Level vials or the two halves of the bubbles line up...

Page 18: ...rence points carefully Poor reference points like those on the heavily worn portion of a machine bed where all the work is performed may mean extra time to bring the machine back into tolerance In oth...

Page 19: ...d one vertical laser plane are needed 1 Mount a target horizontally at the closest reference point to the laser and adjust the target so that it detects the laser 2 Zero the target and move it to the...

Page 20: ...ts described in Measuring Straightness and Measuring Flatness to determine the straightness and flatness of the machine s axes measuring squareness is a simple process To truly measure squareness one...

Page 21: ...e error from the laser not being parallel to the surface Level to Earth Measurements The leveling of machine tools surface plates and different types of bases can be performed using scanning lasers as...

Page 22: ...ny three points on a surface may be used however Hamar Laser recommends the setup illustrated in Figure 12 Figure 12 Three Point Buck in Using One Target recommended setup When performing this procedu...

Page 23: ...ing the Three Point Buck In 1 Center the target in the Near Position With the target in the Near Position zero the target 2 Move the target to Far Position A and tilt the laser beam until the readout...

Page 24: ...he laser beam must be adjusted to be parallel to or coincident with an edge or centerline The process of making that adjustment is called bucking in This section describes two types of buck in methods...

Page 25: ...resents the distance from the laser to the first target L2 repre sents the distance be tween the two targets If L1 is less than one tenth of L2 the normal buck in procedure is used If L1 is greater th...

Page 26: ...ding is to L2 Stated mathematically the ratio is Set L1 Far L2 If L1 L2 and the far reading are known the set can then be determined by the following formula Set Far reading L1 L2 Note This is a simpl...

Page 27: ...two targets When using normal buck in the rule is Zero Near Point Far Buck in the laser beam by zeroing it on the near target and then pointing the laser beam using the appropriate adjustment knobs t...

Page 28: ...eck A C power source Use second A C power supply if available 5 Noisy target A 1519 1520 readings on PDA NOTE On average the user can expect 00002 in ft to 00005 in ft 0 0025 mm m 0 004 mm m of noise...

Page 29: ...low or high on target window it should be near the window It is possible for the laser to activate the auto wake up feature of target but not enough of the beam is hitting the PSD target sensor to get...

Page 30: ...ime Error in software Do not unplug the A 910 from the USB port while still using the program Report to HLI the exact key strokes that created the Runtime Error 12 Software crashes upon loading Ensure...

Page 31: ...27 Appendix B Equipment Drawings L 733 Precision Geometry Laser...

Page 32: ...28 L 743 Ultra Precision Geometry Laser...

Page 33: ...29 The L 732 Dual Scan Precision Roll Alignment Laser...

Page 34: ...30 The L 742 Dual Scan Ultra Precision Roll Alignment Laser...

Page 35: ...ted This makes is very likely that the error in measurement at the point where the target is zeroed is nearly identical to the error in each subsequent measurement because the difference in sensor are...

Page 36: ...ea is 1 micron or less Approx 4 mm of measuring area each vertical line on the graph equals 1 mm Total Error 3 3 microns 2 microns 2 microns 0 000 Each green vertical line equals 1 mm of measuring are...

Page 37: ...the type intended for nail polish removal NOTE When cleaning precision optics even with the best quality optical cleaning tissue use gentle pressure to avoid scratching the surface or damaging the opt...

Reviews: