ORCA-Flash4.0 LT3 Digital CMOS camera C11440-42U40 Instruction manual_Ver.1.1
10
6. NAME AND FUNCTION OF PARTS
•
Do not place the rear panel of the camera, which connectors are
located, to be at the bottom (Do not block ventilation openings.).
①
Lens mount
C11440-42U40 can be attached to C-mount lens or an optics system.
Note
•
The depth of the C-mount is 7 mm. Screwing in the mount too far
can scratch the glass surface.
②
USB 3.1 Gen 1 interface connector [USB 3.1 Gen 1]
This is connected to the USB interface connector on the computer.
③
Trigger input connector [EXT.TRIG]
This is used when the camera is being operated using external synchronization.
Input is TTL or 3.3 V LVCMOS level, and input impedance is 10 k
Ω
.
When an external trigger is input, the trigger is activated at the falling or rising edge of the signal.
(You can choose external trigger polarity between Negative and Positive.)
④
Timing out connector 1,2,3 [TIMING 1,2,3]
This is used when peripheral device(s) require synchronization with the camera.
Output is 3.3 V LVCMOS level, and it is output though
BUS TRANSCEIVER IC SN74LVC541
.
Output impedance is 33
Ω
.
Note
•
Determine termination according to cable length and so on.
⑤
DC power input connector [DC IN]
This is the power supply terminal. Use the accessory AC adapter.
Summary of Contents for ORCA-Flash4.0 LT3