Halma Ocean Optics HL-2000-CAL Installation And Operation Manual Download Page 15

 

 

Chapter 3 

Calibration 

Performing Calibration 

 

Procedure 

Follow the instructions below to calibrate a spectrometer using the HL-2000-CAL: 

1.  Click on the 

Spectrometer

 menu and select 

Calibrate

, then select the channel to calibrate. A 

dialog box opens with the message “Verify lamp was ON for at least 15 minutes for a 
REFERENCE scan.” 

2.  Ensure that nothing is blocking the light path going to your spectrometer. Your setup 

(spectrometer, fiber, and sampling optics) used to obtain a reference should be identical to your 
setup for measuring your sample. Taking a reference spectrum is required before the software can 
calculate absolute irradiance measurements.  

3. Click 

the

 OK 

button. Another dialog box opens with the message “Block light path to 

spectrometer for a DARK scan.”  

4.  Completely block the light path to your spectrometer by inserting an opaque object into the 

lamp’s filter slot (if possible, do not turn off the light source). Taking a dark spectrum is required 
before the software can calculate absolute irradiance measurements.  

5. Click 

the 

OK

 button. The spectral response of your system is now calibrated. If you selected

 

Display calibration info when calculating

 in the

 Configure Spectrometer

 dialog box, you will 

now see the results of the calibration procedure. The top left graph represents the linear 15

th

 order 

polynomial regression. The top right graph represents the dark scan. The bottom graph is the 
calibration curve representing the spectral response of the process. 

The calibration procedure is saved in a file named

 CHO.CAL

 for calibrating the master spectrometer 

channel, 

CH1.CAL

 for calibrating the first slave spectrometer channel, and so on. 

000-10000-060-02-0505 

7

 

Summary of Contents for Ocean Optics HL-2000-CAL

Page 1: ...34698 Phone 727 733 2447 Fax 727 733 3962 8 a m 8 p m Mon Thu 8 a m 6 p m Fri EST Ocean Optics B V Europe Geograaf 24 6921 EW DUIVEN The Netherlands Phone 31 0 26 3190500 Fax 31 0 26 3190505 E mail In...

Page 2: ...2005 Ocean Optics Inc All rights reserved No part of this publication may be reproduced stored in a retrieval system or transmitted by any means electronic mechanical photocopying recording or otherwi...

Page 3: ...anty and create an unsafe operating environment 2 Dangerous voltages are present in this device There are NO user serviceable parts inside 3 Only allow qualified personnel to service this unit 4 Do no...

Page 4: ...Important Safety Notices B 000 10000 060 02 0505...

Page 5: ...ated Documentation iii Upgrades iv Chapter 1 Setup 1 Overview 1 Unpacking the HL 2000 CAL 2 Contents 2 Connecting a Spectrometer to the HL 2000 CAL 3 Configuring the OOIIrrad Software 3 Chapter 2 HL 2...

Page 6: ...Table of Contents ii 000 10000 060 02 0505...

Page 7: ...ing environment specifications as well as other physical details of the product Chapter 3 Calibration Provides instructions for calibrating the spectrometer connected to the HL 2000 CAL Light Source P...

Page 8: ...change or an upgrade to your system To facilitate these changes you must first contact Customer Support and obtain a Return Merchandise Authorization RMA number Please contact an Ocean Optics Applicat...

Page 9: ...e absolute spectral response of your system Note If you have a spectrometer setup that is highly sensitive you may not be able to use the HL 2000 CAL as a calibration source Before using the HL 2000 C...

Page 10: ...a clean laboratory environment see Operating Environment Contents Your DH 2000 CAL package should contain the following DL 2000 CAL unit 1 Switching AC adapter for stabilizing power 1 Power cord for s...

Page 11: ...f the HL 2000 CAL with an Allen wrench Screw the other end of the fiber into the SMA connector of the spectrometer 4 Plug the switching AC adapter the black rectangle shaped box into the back of the H...

Page 12: ...our setup If using a bare fiber enter the fiber s diameter here If using a fiber with a CC 3 cosine corrector enter 3 900 Note Whatever optical setup you wish to use for your application you must also...

Page 13: ...tion Heat Should be situated away from any device that emits excessive heat Object and Liquid Entry Should be positioned so that objects do not fall on top of the unit Additionally ensure that no liqu...

Page 14: ...2 HL 2000 CAL Specifications 6 000 10000 060 02 0505...

Page 15: ...ctrometer for a DARK scan 4 Completely block the light path to your spectrometer by inserting an opaque object into the lamp s filter slot if possible do not turn off the light source Taking a dark sp...

Page 16: ...3 Calibration 8 000 10000 060 02 0505...

Page 17: ...nce iii purpose iii summary iii O OOIIrrad software configuring 3 operating environment 5 P package contents 2 product related documentation iii S setup 1 specifications 5 specifications table 5 spect...

Page 18: ...Index 10 000 10000 060 02 0505...

Reviews: