7-2
Tube Clamp Assembly
P/N 108755-US, Manual revision: AA
T
UBE
C
LAMP
/S
OLENOID
I
NSPECTION
, R
EPLACEMENT
AND
A
DJUSTMENT
Warning: The tube clamp should be treated as a biohazard area. Always wear
rubber gloves when repairing.
Caution:
Always wear an ESD strap when working on the Shaker.
Note: Shakers built after 1996 utilize self-tapping screws. Haemonetics recom-
mends updating Solenoid Assembly mounting screws with the new variety.
1.
With the motor running, gently depress the front of the shroud. The follow-
ing three things should happen.
z
The audible alert will sound momentarily.
z
The red indicator light on the top of the Shaker will illuminate.
z
The solenoid should audibly engage the tube clamp.
2.
If the latter alone fails to occur, or if the motion of the clamp seems slug-
gish, it is possible that the solenoid is providing insufficient force to prop-
erly clamp the tubing. This failure may be due to one of several possible
causes. To locate the problem, it will be necessary to remove the clamp
body from the neck of the Shaker to obtain access to the involved parts.
Figure 7-1, Close-up of Tube Holder and Clamp Body
Nylon Screw
Tube Clamp
Clamp Body
Summary of Contents for SEBRA 1040
Page 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Page 3: ......
Page 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Page 8: ...vi Preface P N 108755 US Manual revision AA...
Page 12: ...x Table of Contents P N 108755 US Manual revision AA...
Page 14: ...xii List of Figures P N 108755 US Manual revision AA...
Page 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Page 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Page 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Page 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Page 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Page 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Page 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Page 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Page 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Page 105: ......