Strain Gauge
5-5
P/N 108755-US, Manual revision: AA
Figure 5-5, Oscilloscope Wave Form
11.
Stop Shaker weighing routine from keypad. Remove power from Shaker.
Detach adjusted assembly from fixture.
Figure 5-6, Weight Sensor Assembly-Current
12.
For those PCBAs at Revision G or higher, after step 5 attach a DVM ground
lead (black) to TP5 and the positive lead (red) to TP2 on the body board.
13.
Apply power to the Shaker. Press the "*" on the keypad to start the Shaker.
14.
With no weight on the knife edge of the strain gage assembly, the DVM
should read between 4.40 and 5.00 VDC.
15.
Hang the strain gage calibration weight (p/n 10401200-00) on the knife
edge, the DVM should read between 6.10 and 7.10 VDC.
16.
Remove the calibration weight, stop the Shaker and remove power. The
weight sensor verification is now complete.
Summary of Contents for SEBRA 1040
Page 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Page 3: ......
Page 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Page 8: ...vi Preface P N 108755 US Manual revision AA...
Page 12: ...x Table of Contents P N 108755 US Manual revision AA...
Page 14: ...xii List of Figures P N 108755 US Manual revision AA...
Page 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Page 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Page 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Page 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Page 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Page 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Page 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Page 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Page 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Page 105: ......