92
Chapter 5, General Operation: Cell Salvage
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
you to resume the previous procedure or to save the previous procedure and
start a new one. If you choose to start a new procedure, the device marks the
previous procedure as complete and performs a self-test. If you instead
chooses to continue the previous procedure, the device instructs you to ensure
all disposables and interlocks are in place.
Additional
Functions
When you end a procedure, the
Record
screen appears, displaying the
procedure record for the current procedure:
Figure 44, Example of the current procedure record
The right side of the screen provides additional actions you can take, including
export the procedure record, view procedure record history for past
procedures, edit the procedure record, and remove air. For more information
on procedure records, see
Removing Air from the RBC Bag
Warning: This process may leave residual air in the RBC bag. Do not
pressure infuse. May cause fatal infusion of air.
To purge any extra air from the RBC bag, you can touch
Remove Air
. A yellow
alert message appears.