162
Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
Event Messages
The following section provides a list of event messages, which appear on the
display screen when an action is required by the user.
Table 16, Event Messages
ID #
Event Message Text
102
Effluent Line Sensor Failure
Explanation:
An issue has been detected with the effluent line sensor. The procedure can be performed, but
the device cannot monitor the effluent from the bowl.
Corrective Action:
To perform the procedure with the user monitoring the effluent:
1. Touch
Continue
.
2. Monitor the effluent from the bowl during processing.
If red cell spillage occurs, manually decrease the pump speed.
OR
To end the procedure:
1. Touch
End Procedure
.
2. If in power-on self-test, power down the device.
If the problem persists, service is required to resolve this issue.
103
Manifold Pressure Sensor Failure
Explanation:
An issue has been detected with the manifold pressure sensor.
Corrective Action:
Restart the device.
If the problem persists, service is required.