Chapter 10, Cleaning and Maintenance
157
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
Air Exhaust Filter
The bottom of the device is equipped with an air exhaust filter. The filter should
be changed periodically, depending upon frequency and conditions of use, to
avoid malfunction resulting from an accumulation of lint and dust in air
passages. If the filter cover is damaged, contact the local Haemonetics
representative for a replacement part.
Warning: Follow local standard operating procedure for blood precautions
when dealing with blood contaminated components.
Follow the steps below to replace the air exhaust filter:
1. Disconnect the device from the power supply.
2. Unlatch the device from the cart.
3. Pull off the black plastic air exhaust filter cover. For easy removal pull from
the corner of the cover.
4. Remove the filter pad and dispose of it in accordance with hospital
guidelines and procedures for biohazardous waste disposal.
5. Insert a new filter pad. Ensure that the textured surface faces the device.
6. Place the black plastic cover back in position and press down on it gently
until it snaps into position.
7. Record the date of maintenance.
Replacing the
Fuses
The fuses are covered by a plastic fuse door, located in the power entry module
at the rear of the device. At the top of the fuse door is a small indentation to
allow the user to open the fuse door. The door is hinged at the bottom.
Follow the steps below to open the fuse door and replace the fuses:
1. Remove the power cord.
2. Use a small screwdriver (1/8” flathead) to open the black plastic fuse door
by placing the screwdriver in the indent at the right side of the door and
levering it open.
3. Using the screwdriver, gently pull out the red fuse holders by placing the
screwdriver under the right side of the holder and levering out the holder.
4. Place new fuses in the fuse holders. Be sure to use appropriately rated
fuses.
5. Replace the fuse holders in the power entry module.
6. Close the fuse door and press gently until it snaps into place.
Inspecting the
Power Cord
Inspect for a frayed or twisted power cord. To order a replacement power cord,
contact your local Haemonetics representative.