Chapter 7, Protocol Settings
127
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
Cell Salvage
Settings
Fat Reduction
(On/Off)
Determines if the device uses a specialized washing sequence designed to
reduce the level of fat in the final product.
On:
The device performs three additional steps during the Wash phase
that isolate the fat and remove it from the bowl.
Off:
The device performs a normal Cell Salvage Wash phase.
Sequestration
Parameter
Values
Default
PPP Return Plasma
No/Yes
No
PRP Milk
No/Yes
No
Parameter
225 mL Bowl
125 mL Bowl
70 mL Bowl
Min
Max
Step
Fill Pump Rate (mL/min)
60
60
N/A
10
250
10
Fill Centrifuge Speed
(RPM)
5650
5650
N/A
2050
5650
50
PRP Collection Pump
Rate (mL/min)
20
20
N/A
10
250
10
PRP Collection
Centrifuge Speed (RPM)
2450
2450
N/A
2050
5650
50
Empty Pump Rate
f
(mL/min)
500/400/100
300/150/100
N/A
25
1000
25
PRP Extend Volume (mL)
10
10
N/A
0
50
1
PRP Milk Volume (mL)
3
3
N/A
1
50
1
PRP Milk Delay (seconds)
10
10
N/A
1
30
1
a. Pump regulation is not applicable for the 70 mL bowl.
b. The volume accounting parameter is only available if the empty destination is set to the red line.
c. If pump regulation is on, the device optimizes the quality of the final product by adjusting the pump
speed during the Fill phase from 150 - 500 mL/min (150-225 mL/min for a 125 mL bowl), during the
Wash phase from 150-450 mL/min (150-200 mL/min for a 125 mL bowl), during the Concentrate
phase from 125-150 mL/min (75-125 mL/min for a 125 mL bowl) and during a Wash phase after a
Concentrate phase from 100-200 mL/min (75-175 mL/min for a 125 mL bowl) based upon the effluent
line sensor readings.
d. Minimum wash volume for the 70 mL bowl is 300 mL
e. Step is 100 mL for the 70 mL bowl.
f. If Smart Empty is on, the device reduces the pump speed during the Empty/Return phase in
preprogrammed increments to minimize the number of RBCs remaining in the bowl at the end of the
Empty/Return phase.
Table 11, Default Settings