Chapter 6, General Operation: Sequestration
107
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
Figure 53, Hanging the waste bag
Installing the
Blood Bag
Adaptor
Harness
Follow the steps below to install the blood bag adaptor set:
1. Close the ratchet clamps on the yellow saline lines of the processing set.
2. Connect the red line from the processing set to the red line connection on
the adaptor set.
3. Connect the reservoir drain port connection on the adaptor set to the
reservoir drain port.
Note: If a reservoir is not going to be used for Cell Salvage following the
Sequestration procedure, the adaptor harness does not need to be connected
to a reservoir. In this case, the slide clamp on the port that would connect to the
adaptor harness should be closed.
1.
2.
3.
1. Effluent tubing
connection
2. Waste bag pins
3. Waste bag drain port