Chapter 6, General Operation: Sequestration
103
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
Installing the Sequestration Disposables
Inspecting the
Disposable Sets
Always inspect disposable sets while removing them from the packaging.
1. Read the labeling on the disposable set to ensure it is the correct set for
the current procedure.
2. Ensure there are no kinks or twists in the tubing that could restrict the flow
of fluid.
3. Check that there are no missing caps or open connections.
4. Verify that there are no visible defects or particulate within the set.
Loading the
Reservoir and
Vacuum Line
The Sequestration protocol is only available prior to starting the first Fill phase
in the Cell Salvage procedure.
To prepare for the Sequestration protocol, load the reservoir and vacuum line:
1. Place the reservoir in the reservoir weigher so that the three filtered inlet
ports face the tubing support.
Note: The reservoir weigher should be no higher than 183 cm (72 in.) from the
floor.
2. Close the slide clamp on the reservoir drain port. Do not load the A&A line
at this time. Leave the port capped until ready to use.
3.
If using the Cell Saver Elite internal suction,
connect the filtered vacuum
line to both the vacuum port on the back of the device and to the vacuum
inlet port on the reservoir.
4.
If using external suction,
connect the external vacuum to the vacuum inlet
port on the reservoir.
Installing the
Processing Set
Selecting the Bowl Size
1. If you have not already selected the Sequestration protocol, touch
(
Menu
) and select
Sequestration
from the drop-down list. The device
displays the Sequestration
Bowl Selection
screen.
2. From the
Bowl Selection Screen
, scan a processing set using the barcode
reader underneath the touch screen display or select the appropriate bowl
size on the touch screen. The
Processing
screen appears.
Note: Sequestration is not available with the 70mL bowl set.
3. Extend the tub holder located on the right side of the cart.
4. Place the tub in the holder so that the top of the bowl faces the back of the
device.