Manual Operation
P/N 53063-50, Manual Revision: B
6-8
EMERGENCY
Option
An emergency option is available from the MANUAL mode. This option is
described in detail in Chapter 5. The EMERGENCY option is not available for use
with the 70 ml bowl. To enter the EMERGENCY option from the MANUAL mode,
press the MODE key twice. The operator will then be prompted to confirm this
choice by pressing the YES key within 5 seconds. If the YES key is not pressed
within 5 seconds, or the NO key is pressed, the CS5+ will revert to the
AUTOMATIC mode.
The CS5+ is programmed with preset processing parameters that can be modified
by the operator while in MANUAL operation. These parameters can all be
modified from the STANDBY state or some parameters can be changed at various
points in the processing cycle.
To modify the preset parameters the operator should:
1. Press the MODIFY key and the following will be displayed:
2. Use the SELECT key to move from parameter to parameter. Use the UP/
DOWN arrows to adjust the values up or down or to toggle between
selections.
3. Once done with changing the parameters press MODIFY to return to
STANDBY.
Note: WHILE IN MANUAL MODE, ALL modifications to the preset parameters
will be retained from procedure to procedure until reprogrammed by the
operator.
CELL SAVER
Press ARROW keys to change values
Press SELECT to advance highlight
MANUAL
Press MODIFY when finished
SETUP
FILL rate
WASH rate:
EMPTY rate:
Centrifuge Speed:
Waste Bag Weigher:
ALARM Sound:
500 ml/min
500 ml/min
500 ml/min
5650 rpm
ON
ON