P/N 53063-50, Manual Revision: B
Automatic Operation: Cell Salvage
5-5
The operator can begin a FILL mode at anytime during Standby by pressing the
START key. But at this point, the operator has other options, depending upon
which keys are lit. The operator can:
a. Press the backlit WASH key to advance to the wash mode and wash a
partially filled bowl (see previous discussion in Chapter 2; twice the standard
wash volume should be used).
b. Press the backlit RETURN key to move the contents of the bowl back into the
reservoir for future processing.
Note: When the bowl size is detected, the bowl size appears inside the bowl icon
on the screen.
5. When the optical sensor detects the bowl is full of cells, or when the
operator presses the backlit WASH key, the system will enter the wash mode
and bring saline solution into the bowl. The saline is pumped into the bowl
and flows through the red blood cells, washing debris, free HgB, heparin or
other anticoagulant, activated clotting factors, etc., into the waste bag.
The following display will be shown while the system is in the WASH mode:
Pressing YES will double the program wash volume for this wash cycle only. This
prompt will be available for the entire wash sequence until the minimum wash
volume is reached. When the double wash volume WASH mode is complete, the
machine will revert back to the original programmed wash volume.
If minimum wash volume is manually modified by the operator, the volume
selected will be used throughout the procedure instead of the preprogrammed
value. If the wash volume is 1000 ml or greater for a 225 ml bowl (750 ml or
greater for a 125 ml bowl or 300 ml or greater for the 70 ml bowl), this
modification will be permanent until changed by the operator.
225
ML
WASH CELLS
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
Wash Volume: 150/1000 ml
Press MODE twice to enter EMERGENCY option
AUTOMATIC
800 ml
150 ml
0 ml
0
Pump Rate:
Press MODIFY to change minimum wash volume
500 ml/min
NaCl
Press YES to double the programmed wash volume