P/N 53063-50, Manual Revision: B
Setup Procedures for Cell Salvage
4-3
S
ETUP
Power On
Procedure
When ready to begin processing, or to install disposables using the HELP
function, press the POWER ON switch located on the right side of the CS5+.
If the machine has been restarted within the past two hours, the CS5+ gives you
the option of saving information, as shown on the following screen:
Figure 4-1: Power OFF recovery
Press YES if you would like to retain the data from before the unit was turned off.
Press NO to start a new procedure.
The unit then goes into self-test to insure that all the functions are operating
before the unit is allowed to be used in a procedure. A SELF-TEST screen will be
displayed indicating the stage of self-test completion:
Figure 4-2: Self-Test Display
X.X indicates the current revision of software installed on the machine.
POWER OFF RECOVERY
Power has been off for 12 minute(s).
Continue the previous procedure?
Press YES to keep previous data
Press NO to begin a new procedure
!
HAEMONETICS
CORPORATION
CELL SAVER
Self-Test in Progress
30%
CS5+-X.X