Maintenance
P/N 53063-50, Manual Revision: B
8-4
Cleaning the
Effluent Line
Sensor
The effluent line sensor should be cleaned by flossing with a water dampened
lint-free cloth followed by flossing with a dry lint-free cloth.
Air Detector
Cleaning
The air detector should be cleaned by flossing with a water dampened cloth
followed by a dry lint-free cloth.
Cleaning the
Tubing Valves
The tubing valves and clamped line sensor may be cleaned by wiping with a mild
detergent-dampened lint-free cloth.
Blood Pump
Cleaning
The blood pump should be cleaned after a spill to keep the rotating parts moving
freely. The pump rotor may be lifted out by holding the rotor stationary and
unscrewing the cover. Clean the area under the pump rotor and pump arm until
all moving parts can rotate or slide freely. Secure the pump rotor back into its
original position. Wipe inside the pump assembly with mild detergent-dampened
cloth. Rinse the pump under the faucet with warm water. If required, pump rotor
can be cleaned in an ultrasonic cleaner. Dry thoroughly.
Cleaning After a
Spill
The exterior of the CS5+, including the control panel, should be cleaned with a
mild detergent or disinfectant at regular intervals as well as any time a spill has
occurred. Do not use 100% bleach directly on the machine; a bleach/water
solution or some other disinfectant may be indicated. The air filter should also be
cleaned as described in the next section.
Note: If the mechanical chuck becomes contaminated with blood, call
Haemonetics hotline at (800) 433-3431 for further assistance. If blood has
entered the finger-locking mechanism, the mechanical chuck must be cleaned by
a trained biomed or a haemonetics field service engineer.
Cleaning the Air
Filter
The CS5+ is equipped with a rear panel air filter for incoming cooling air. Correct
filter maintenance will avoid malfunction resulting from an accumulation of lint
and dust in air passages. The filter should be washed periodically, depending
upon frequency and conditions of use.
Supplies Needed
The only supply needed is warm running water.