Sequestering on the Cell Saver 5+ System
P/N 53063-50, Manual Revision: B
7-14
Sequester Setup
Terminology
Fill Bowl DRAW Rate:
The programmed pump speed in milliliters per minute the
pump runs while the bowl is filling with whole blood and during the PPP
collection phase.
Fill Bowl Centrifuge Speed:
The programmed centrifuge speed in revolutions per
minute the centrifuge turns while the bowl is filling with whole blood and during
the PPP collection phase.
PRP Collection DRAW Rate:
The programmed pump speed in milliliters per
minute the pump runs during the PRP collection phase.
PRP Collect Centrifuge Speed:
The programmed centrifuge speed in revolutions
per minute the centrifuge turns during the PRP collection phase.
PRP Extend Collection Volume:
The volume in milliliters to pump after the line
sensor detects RBCs during the PRP collection phase. Once this volume is
pumped, the PRP collection is considered complete.
Return Plasma?
This question is asked to determine which clamps to prompt the
operator to open and close at the beginning of the EMPTY phase and to account
for volume returned. If the operator selects YES, the PPP is returned to the bowl
and then to the reinfusion bag along with the RBCs.
Milk PRP Collection Layer?
This question is asked to determine if the process of
starting and stopping the PRP collection phase is desired as determined by the
programmed parameters “Milk Vol” and “Milk Delay.”
Milk Vol:
The volume to pump in milliliters during the PRP collection until the
pump stops and the Milk Delay time starts.
SEQUESTER SETUP - ASSISTED SEQUESTER
PARAMETER
Default
Max
Min
↑↓
Assisted Sequester Mode:
OFF
ON
OFF
Toggle
Fill Bowl DRAW Rate:
60
250
10
10
Fill Bowl Centrifuge Speed:
4750
5650
2050
100
PRP Collection DRAW Rate:
60
250
10
10
PRP Collect Centrifuge Speed:
4750
5650
2050
10
PRP Extend Collection Volume:
30
300
0
5
Return Plasma?
NO
YES
NO
Toggle
Milk PRP Collection Layer?
NO
YES
NO
Toggle
Milk Vol:
3
25
1
1
Milk Delay:
10
30
1
1