P/N 53063-50, Manual Revision: B
Sequestering on the Cell Saver 5+ System
7-5
2. Press MODIFY to see the SETUP screen.
Use the SELECT key to highlight parameters and use the ARROW keys to change
values.
3. Press MODIFY to see the SEQUESTER SETUP screen.
Use the SELECT key to highlight parameters and use the ARROW keys to change
values.
The preset default parameters for the manual sequester
mode are as follows:
Assisted Sequester Mode:
OFF
Fill Bowl DRAW Rate:
60 ml/min
Fill Bowl Centrifuge Speed:
4750 rpm
Return Plasma?
NO
Press ARROW keys to change values
Press SELECT to advance highlight
Min Wash Vol. - 225:
-125:
Reservoir Level:
Resume at Level:
Reservoir Level Sensor:
Pump Regulation:
Empty RBCs to:
AUTOMATIC
Press MODIFY when finished
NO WASH
NO WASH Option:
Protocol:
OFF
Cell Saver
1000 ml
750 ml
800 ml
400 ml
ON
ON
Blue Line
CELL SAVER SETUP
SEQUESTER SETUP
Press ARROW keys to change values
Press SELECT to advance highlight
Assisted Sequester Mode:
Fill Bowl DRAW Rate:
MANUAL
Press MODIFY when finished
OFF
60
rpm
Fill Bowl Centrifuge Speed:
Return Plasma?
4750
NO
rpm