P/N 53063-30, Manual revision: B
1-11
Presenting the Cell Saver 5+ System
Note: Haemonetics will regulate the proper voltage setting upon installa-
tion. The power source used must be properly grounded.
Caution: The Cell Saver 5+ device must be operated in an environment
compatible to the requirements of the IEC 60601-1-2:2001 Standard, Elec-
tromagnetic compatibility.
Mobile RF communication equipment not approved by Haemonetics and
portable communication equipment can affect the Cell Saver 5+ device.
Any accessories and cables not approved by Haemonetics used in conjunc-
tion with the device may increase hazards and influence compatibility with
EMC requirements. Therefore, non-approved accessories and cables must
not be used.
In addition, the Cell Saver 5+ device and accessories must not be placed
directly adjacent to, or top of other equipment, unless specifically approved
by Haemonetics.
Fluid management systems
The device is equipped with a pole for hanging the saline solution, reservoir,
reinfusion bag, anticoagulant solution and transfer packs. The pole may be
raised and lowered by loosening the lower knob. The top portion can be
extended by releasing the upper knob. The movable pole hooks are used for
hanging the saline and reinfusion bags.
The waste bag weigher provides three hooks on the front of the device for
hanging and monitoring the level in the waste bag.
Processing speeds
!
Typical processing time: 5 minutes.
!
Centrifuge speed:
!
For the Latham bowls:
5650 rpm, programmable from 2050 rpm
to 5650 rpm in 100 rpm increments.
!
For the 70mL Bowl:
7000 rpm, programmable from 2050 rpm to
7000 rpm in 100 rpm increments.
!
Pump speed:
!
For the Latham bowls:
programmable from 25 to 1000 ml/min in
25 ml/min increments.
!
For the 70mL Bowl:
programmable from 25 to 300 ml/min in 25 ml/
min increments.
!
May be temporarily reset with Pump Control arrow keys.