BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
46
influence compatibility with EMC requirements. Non-approved accessories
and cables must
not
be used.
•
A power cord is supplied with the BloodTrack Kiosk V5. Do not replace the
power cord with a substitute. If necessary, contact Haemonetics Technical
Support
—see
Appendix C
—for a replacement. Use
only
Haemonetics-approved peripherals, cabling and connections when installing
the BloodTrack Kiosk V5.
CAUTIONS
•
The electrical power outlet used for the power cord for the BloodTrack
Kiosk V5 should be easily accessible to allow it to be disconnected in order
to remove mains power. The BloodTrack Kiosk V5 should not be installed
with a permanent electronic supply feed.
•
The power outlet used to connect to the BloodTrack Kiosk V5 may come
from the optional UPS. Therefore the only certain method to disconnect the
mains supply from the BloodTrack Kiosk V5 is to disconnect the IEC C13
cordset from the C14 inlet on the chassis. The UPS may still provide mains
power to the BloodTrack Kiosk V5 if mains power input to the UPS itself is
disconnected.
External cabling such as network cables, should run inside suitable trunking, where
possible, to provide a neat installation.
8.3
Storage Location Door Locks
CAUTION
When the BloodTrack Kiosk V5 relay door lock interface is used to drive a
magnetic lock, a diode
must
be fitted as per
Section 8.3.7
, otherwise the
BloodTrack Kiosk V5 may suffer from unpredictable behavior.
8.3.1 Storage Location Interfaces
The BloodTrack Kiosk V5 has three ports available for interfacing to storage
locations. These are available in two configurations:
1. One relay output and two serial port outputs.
2. Three relay outputs.
8.3.2 Interface Connector
The BloodTrack Kiosk V5 uses a detachable terminal block connector to provide
either the relay output or the serial port output depending on the port configuration.