146
Disassembly
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
Vacuum Display
1. Remove the membrane panel assembly from the top cover assembly.
2. Disconnect the connectors from the vacuum display.
3. Place the membrane panel on a flat level surface and remove the (4)
Phillips head screws securing the vacuum display to the top cover
mounting plate. Note that there may be (4) nylon washers located under
the vacuum display that may otherwise go unnoticed. The flat, level
surface will help to keep the washers localized.
4. Carefully lift the vacuum display up and away from the top cover
mounting plate.
Note: If the vacuum display was elevated by nylon washers, be sure to replace
them before installing the vacuum display.
Note: When replacing the vacuum display, call the Haemonetics Hotline and
verify the correct parts to order. Depending on the present display
configuration, it may be necessary to add a shim between the vacuum display
and the membrane panel. The newer vacuum displays are slightly thinner than
the older versions. This shim can be ordered from Haemonetics.
Membrane
Panel
1. Remove the membrane panel assembly from the top cover assembly,
and remove the display distribution PCB from the top cover mounting
plate.
2. Remove the (6) perimeter 1/4” nuts securing the top cover mounting
plate to the top cover panel fascia.
3. Remove the (1) 1/4” nut securing the membrane panel ground strap to
the membrane panel mounting lug and pull the ground strap off the lug.
4. Remove the (7) perimeter and (1) central 1/4” nuts, and the (1) 5/16” nut
securing the membrane panel to the top cover mounting plate.
5. Pull the top cover mounting plate away from the top cover panel fascia.
6. Remove the nylon spacer from the central membrane panel mounting
lug.
7. Using considerable force, peel the membrane panel edges away from
the top cover panel fascia and push the membrane panel through the top
cover panel fascia.
The membrane panel cannot be reused!