Connection diagram

Sensor B does  not  exist   

in case of single-side  

measurement

Different connection 

in case of explosion 

proof application

The four terminal blocks with five terminals 

each can be seperatly plugged in, simplifying 

trouble shooting in case of faults.

Recommendation: Connect the PE-terminal No. 5 of the amplifier with a 

short cable with one of the separate PE-terminals on the DIN rail.

PE terminal

PE terminal

yellow

white brown

green

yellow

white

brown

green

V

1

Output signal of full bridge  

strain gauge

V

2

Direct voltage output

V

3

Filtered voltage output

V

4

Excitation voltage to the full 

 

bridge strain gauge in the sensors

V

5

Supply voltage 24 V DC

I

1

Current output (option C and N)

For Zero point adjustment use the  

voltage output (V

2

)

HAEHNE

  Elektronische Messgeräte GmbH · Heinrich-Hertz-Str. 29 · D-40699 Erkrath Germany · Telefon 0211/9 25 91-0 · Fax 0211/9 25 91-20

http://www.haehne.de 

Email: [email protected]

Digital Measuring Amplifier DMA2

Technical Information

Sensor cable lead color 

with connection variant:

 

N3, S, S3, S4, T 

 V

1

: Signal voltage

 V

4

: Supply voltage

  

Sensor cable lead color 

with connection variant:

 

N1, N2, S1, S2 

  