Appendix B Frequently asked questions
Table 24 Application files
Question
Cause
Procedure
Why can't I load any applications
Wrong file placement on the
USB storage device
Make sure that the file path on the USB storage
device is correct
The user is not in the home
screen
Push the
Home
key to make sure that you are in
the home screen before connecting the USB
storage device
Why is the instrument not recognizing
the applications USB storage device,
even though I have removed and
reconnected the device and I see the
USB icon in the header bar
Reconnecting the device has
been done too quickly
Remove the device and wait until the icon
disappears from the header bar, then push the
Home
key to make sure that you are in the home
screen before reconnecting the device
Table 25 Mechanical
Question
Cause
Procedure
How can I remove an air bubble that is trapped
on top of the syringe
Air bubbles are trapped in
the syringe
Select the
Purge
option on the instrument to
purge the syringe of air bubbles (refer to
Prepare the instrument for measurement
on page 27)
Why is titrant leaking along the sealing above
the electrovalve
The tubes have not been
correctly installed
Turn the tube connector until a "click" is
heard
Why is titrant leaking along the sealing below
the electrovalve
The syringe has not been
installed tight enough
Make sure that the syringe is installed
correctly
The electrovalve is leaking
Change the electrovalve as described in
Replace the syringe electro-valve block
on page 40
On my two pump instrument, using the
Pump
activation
option, why is the reagent not being
delivered into the beaker
The procedure is not being
done on the correct pump
Select the
Pump activation
option on the
instrument, and
Toggle
the pumps to make
sure that the correct pump is being used
The active pump is not the
correct pump
Change to expert mode and change the
name of the pump
The pump is defective
Change the pump as described in
on page 42
Table 26 Password
Question
Procedure
How can I retrieve my password to access expert mode
Contact technical support
63
Summary of Contents for TitraLab KF1000 Series
Page 1: ...DOC022 52 93082 TitraLab KF1000 series workstations User Manual 03 2019 Edition 9...
Page 2: ......
Page 6: ...Specifications 4...
Page 12: ...General information 10...
Page 20: ...Installation 18...
Page 30: ...Startup 28...
Page 40: ...Standard operations 38...
Page 54: ...Troubleshooting 52...
Page 60: ...Replacement parts and accessories 58...
Page 64: ...Change application parameters 62...
Page 66: ...Frequently asked questions 64...
Page 67: ......