38
SECTION 3,
continued
4. Press the
ENTER
key three times to skip to the display
showing
100%
.
5. Press the
ENTER
key. The stabilizing icon will appear while
the meter completes the calibration.
6. When the calibration is complete, the meter will return to the
Read Mode. Press the
EXIT
key during the calibration
sequence to back out of the calibration routine, one screen
one-at-a-time, without completing a calibration.
3.3.7
Calibration Review
To review the last calibration:
1. Press the
REVIEW
key on the keypad. The date and year of the
last calibration will show.
2. Press the
TIME
key on the keypad to view the time of the
last calibration.
3. Press the
UP ARROW
. The dissolved oxygen concentration of
calibration will show.
4. Press the
CONC %
key to view the % saturation and mg/L
values of calibration.
5. Press the
UP ARROW
key. The barometric pressure entry of
calibration will show.
6. Press the
UP ARROW
key. The altitude entry of calibration
will show.
7. Press the
UP ARROW
key. The salinity entry of calibration
will show. Press the
EXIT
key to leave the calibration review.