Table 1 Changing Meter Setup Options

Setup 

How To Get There

Description of Option

1

Turn the meter on and press 
SETUP/CE. Press READ/ENTER.

Display continuous dissolved oxygen readings or lock the 
reading when the value stabilizes. READ/ENTER

 

toggles 

between the options. Press EXIT

 

when selection is 

complete.

2

Turn the meter on and press 
SETUP/CE

Press the up arrow 

once. 

Use Metric (°C, mm Hg, meters) or English (°F, in. Hg, feet) 
units by showing °C or °F. READ/ENTER

 

toggles between 

the options. Press EXIT

 

when selection is complete.

3

Turn the meter on and press 
SETUP/CE

Press the up arrow 

twice. 

Display D.O. concentrations in 

0.0 mg/L

 or

 0.00 mg/L

 

resolution. READ/ENTER

 

toggles between the options. 

Press EXIT

 

when selection is complete.

4

Turn the meter on and press 
SETUP/CE

Press the up arrow 

three times. Press READ/ENTER.

Adjust the displayed D.O. concentration based on sample 
salinity. Use the number keys to enter a salinity value 
ranging from 0–42. Press READ/ENTER and then

 

EXIT

 

when selection is complete.

5

Turn the meter on and press 
SETUP/CE

Press the up arrow 

four times. Press READ/ENTER.

Use the number keys to enter the barometric pressure. 
Press READ/ENTER and then EXIT

 

when selection is 

complete. See 

Table 2 below.

6

Turn the meter on and press 
SETUP/CE

Press the up arrow 

five times. Press READ/ENTER.

Use the number keys to enter the altitude. 
Press READ/ENTER and then

 

EXIT

 

when selection is 

complete. See 

Table 2 below.

7

Turn the meter on and press 
SETUP/CE

Press the up arrow six 

times. Press READ/ENTER.

Use the number keys to enter the time. 
Press READ/ENTER and then

 

EXIT

 

when selection is 

complete.

8

Turn the meter on and press 
SETUP/CE

Press the up arrow 

seven times. Press READ/ENTER.

Use the number keys to enter the date. 
Press READ/ENTER and then

 

EXIT

 

when selection is 

complete.

9

Turn the meter on and press 
SETUP/CE

Press the up arrow 

eight times. Press READ/ENTER.

Use the number keys to enter the year. 
Press READ/ENTER and then

 

EXIT

 

when selection is 

complete.

10

Turn the meter on and press 
SETUP/CE

Press the up arrow 

nine times. Press READ/ENTER.

Use the arrow keys to select time intervals for automatic 
data transfer. Press READ/ENTER and then

 

EXIT

 

when 

selection is complete.

Table 2 Adjusting Barometric Pressure and Altitude

Enter a new barometric pressure when the barometric pressure or the instrument altitude changes.

Using Sea Level Equivalent

Using True Barometric Pressure

1.

Obtain the sea level equivalent barometric 
pressure from TV, radio, or a local airport.

1.

Obtain the true barometric pressure from a 
nearby mercury barometer or use 

Table 11 on 

page

74 of the manual.

2.

Enter this value into the meter in Setup 5. 

2.

Enter this value into the meter in Setup 5 or 
during calibration.

3.

Enter the local altitude in Setup 6.

3.

Enter the altitude as 0 feet or meters.