Startup
Turn the meter on and off
N O T I C E
Make sure that the probe is connected to the meter before the meter is turned on.
Push and hold to turn on or turn off the meter. If the meter does not
turn on, make sure that the batteries are properly installed.
Note: The meter is set to automatically power off after 5 minutes of inactivity to
maximize the battery life.
Operation
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.
Setup for pH
The meter is ready for pH or ORP calibration without additional setup.
Setup for dissolved oxygen
Before calibration or measurement for dissolved oxygen, the probe must
be polarized and the settings for atmospheric pressure and salinity must
be entered.
Polarize the electrode
If the probe or batteries are disconnected, connect the probe or install
the batteries and wait for polarization:
Disconnect time
Polarization time
< 5 minutes
10 minutes
5 to 15 minutes
45 minutes
> 15 minutes
6 hours
Change the settings
The settings for dissolved oxygen measurements can be changed when
a dissolved oxygen probe is used. Push
to view the current settings.
1.
Push
. The first setting is shown.
2.
Quickly use the arrow keys to change the value. Wait for the next
setting to be shown and change any additional values:
Option
Description
Atmospheric pressure in mbar (default: 1013 mbar)
SAL
Salinity in g/L NaCl (default: 0 g/L)
Aut
: automatic (only in model MM156)
SEn
Select the DO sensor model:
5120
or
5130
Note: The salinity value can be found from a conductivity measurement. Refer
to
on page 17.
Setup for conductivity
The settings for conductivity measurements can be changed when a
conductivity probe is used. Push
to view the current settings.
1.
Push
. The first setting is shown.
2.
Quickly use the arrow keys to change the value. Wait for the next
setting to be shown and change any additional values:
Option Description
CEL
Electrode type: platinum (default) or titanium.
Note: This option is not available on all meters.
CAL
The number of standards for conductivity calibration. Select one
standard (default). Select the standard nearest to the expected
measurement range.
tC
Temperature compensation: 0 to 9.99%/°C (default: 2%/°C)
tr
Reference temperature: 20 or 25 °C (default: 25 °C)
F
TDS calculation factor: 0.01 to 4.44 (default: 0.64)
10
English
Summary of Contents for sensION+ EC5
Page 194: ...TDS 0 00 o 4 44 0 64 5 CE 194...
Page 196: ...1 2 3 4 5 NiMH NiMH 1 2 V 1 5 V 2300 mAH NiMH 90 NiMH AA 5 Display Options 196...
Page 197: ...1 IP67 1 1 2 2 2 197...
Page 198: ...3 1 4 2 5 3 3 4 pH1 1 pH ORP 4 2 5 3 198...
Page 199: ...5 DO6 1 DO 4 2 5 3 6 MM110 1 pH 4 2 5 3 6 ORP 199...
Page 200: ...7 MM150 1 5 2 pH 6 3 7 TDS ORP 4 8 8 MM156 1 5 2 pH 6 3 7 DO 4 8 200...
Page 201: ...9 EC5 1 4 2 5 3 6 TDS 5 MSDS pH pH ORP 201...
Page 203: ...1 2 3 4 5 6 pH 7 pH 8 9 10 11 1 2 pH ORP 1 2 3 pH ORP mV pH ORP 1 2 3 OFF 203...
Page 205: ...5 1 2 OFF 3 OFF 1 2 2 3 3 2 205...
Page 206: ...10 10 E1 206...
Page 207: ...E2 250 nA E2 pH 70 105 E2 30 E3 0 05 1 207...
Page 208: ...E3 58 mV E4 E4 TC 0 E5 E6 ErA MM156 50 g L 208...
Page 259: ...0 00 4 44 0 64 5 CE 259...
Page 261: ...2 3 4 5 NiMH NiMH 1 2 1 5 NiMH 2300 90 NiMH 5 261...
Page 262: ...1 IP67 1 1 2 2 2 3 1 4 2 5 3 262...
Page 263: ...3 4 pH1 1 pH 4 2 5 3 5 DO6 1 4 2 5 3 263...
Page 264: ...6 MM110 1 pH 4 2 5 3 6 7 MM150 1 5 2 pH 6 3 7 4 8 264...
Page 265: ...8 MM156 1 5 2 pH 6 3 7 4 8 9 EC5 1 4 2 5 3 6 265...
Page 266: ...5 MSDS pH pH 5 10 5 15 45 15 6 1 2 1013 SAL NaCl 0 Aut MM156 SEn DO 5120 5130 3 274 266...
Page 269: ...5 6 85 C 2 1 2 C F 5 1 2 OFF no 3 OFF 1 2 2 3 3 2 269...
Page 270: ...10 10 E1 270...
Page 271: ...E2 250 E2 pH 70 105 E2 30 E3 0 05 1 271...
Page 272: ...E3 58 E4 E4 TC 0 TC E5 E6 ErA MM156 50 TC 272...
Page 275: ...3 NaCl 33 6 20 2 41 2 25 4 48 9 30 6 56 3 35 8 62 2 40 0 69 4 45 3 75 7 50 0 275...
Page 338: ...TDS 0 00 4 44 0 64 5 CE 338...
Page 340: ...3 4 5 NiMH NiMH 1 2 V 1 5 V 2300 mAH NiMH 90 NiMH 5 340...
Page 341: ...1 IP67 1 1 2 2 2 3 1 4 2 5 ON OFF 3 341...
Page 342: ...3 Measure Calibration Standby 4 pH1 1 pH ORP 4 2 5 3 5 DO6 1 DO 4 2 5 3 342...
Page 343: ...6 MM110 1 pH 4 2 5 3 6 ORP 7 MM150 1 5 2 pH 6 3 7 TDS ORP 4 8 343...
Page 344: ...8 MM156 1 5 2 pH 6 3 7 4 8 9 EC5 1 4 2 5 3 6 TDS 344...
Page 348: ...5 6 85 C 2 1 2 C F 5 1 2 OFF no 3 OFF 1 2 2 3 3 2 348...
Page 349: ...10 10 E 1 349...
Page 350: ...E 2 250 mA E 2 pH 70 105 E2 30 E3 0 05 cm 1 350...
Page 351: ...E3 58 mV E4 E4 TC 0 TC E5 E6 ErA MM156 50 g L TC 351...
Page 354: ...3 mS cm g L NaCl 56 3 35 8 62 2 40 0 69 4 45 3 75 7 50 0 354...
Page 386: ......
Page 387: ......