background image

51970-88

DISSOLVED OXYGEN PROBE

Introduction

This instruction sheet provides information on the Hach Dissolved 
Oxygen Probe for use with the sens

ion

6 Dissolved Oxygen Meter. 

The Dissolved Oxygen Probe is a Clark-type amperometric sensor used 
to measure dissolved oxygen in aqueous solutions. It consists of an 
anode/cathode electrode system and potassium chloride-based 
electrolyte, separated from the sample by a replaceable oxygen-
permeable Teflon membrane. A built-in thermistor provides automatic 
temperature compensation when using the sens

ion

6 Dissolved 

Oxygen Meter.

Specifications

Dissolved Oxygen 
Probe
1-meter cable

Dissolved Oxygen 
Probe
3-meter cable

Dissolved Oxygen 
Probe
15-meter cable

Warranty

1 year

1 year

1 year

Concentration Range

0 - 20 mg/L or 
0 - 200% saturation 

0 - 20 mg/L or 
0 - 200% saturation 

0 - 20 mg/L or
 0 - 200% saturation 

Accuracy

± 1.0% Full Scale

± 1.0% Full Scale

± 1.0% Full Scale

Temperature Range

0 - 50 °C

0 - 50 °C

0 - 50 °C

Response Time

99% of full value in 
1 minute

99% of full value in 
1 minute

99% of full value in 
1 minute

Minimum Sample 
Flow

20 cm/sec

20 cm/sec

20 cm/sec

Drift

< 1%/day

< 1%/day

< 1%/day

Length

150 mm

150 mm

150 mm

Diameter

12 mm

12 mm

12 mm

Connector

5-pin circular

5-pin circular

5-pin circular

Weight Accessory

Optional

Optional

Included