Hach Polymetron 9582sc User Manual Download Page 1

DOC023.53.93067

Hach 9582sc Dissolved

Oxygen

02/2020, Edition 5

User Manual

Summary of Contents for Polymetron 9582sc

Page 1: ...DOC023 53 93067 Hach 9582sc Dissolved Oxygen 02 2020 Edition 5 User Manual...

Page 2: ......

Page 3: ...iguration 10 7 2 Configure the sensor 10 7 3 Calibrate the sensor 11 7 3 1 About sensor calibration 11 7 3 2 Temperature calibration 11 7 3 3 Zero calibration 11 7 3 4 Calibration in air 12 7 3 5 Cali...

Page 4: ...Table of Contents 2...

Page 5: ...pressure Atmospheric pressure Relative humidity 0 90 Certifications EN 61326 1 2006 EN 61010 1 2010 Sensor module Specification Details Measuring range 0 2 ppm Sensitivity 0 5 ppb Repeatability 0 5 p...

Page 6: ...if not avoided could result in death or serious injury C A U T I O N Indicates a potentially hazardous situation that may result in minor or moderate injury N O T I C E Indicates a situation which if...

Page 7: ...t transfer liquid 3 3 Product components Make sure that all components have been received If any items are missing or damaged contact the manufacturer or a sales representative immediately Components...

Page 8: ...il there is resistance Do not force the electrode 4 Install and tighten the holding nut by hand 5 Install the gasket and filling screw to prevent electrolyte leakage or external contamination from the...

Page 9: ...connector not provided 3 Assembled sensor 7 M4 screws not provided 4 Gasket 8 1 4 NPT sample output connector not provided 4 4 Install the controller Refer to the controller documentation to see mount...

Page 10: ...c components can be damaged by static electricity resulting in degraded performance or eventual failure Refer to the illustrated steps that follow and Table 1 Be sure to connect all sensor ground shie...

Page 11: ...English 9...

Page 12: ...on and push enter To enter numbers characters or punctuation push and hold the up or down arrow keys Push the right arrow key to advance to the next space Option Description EDIT NAME Changes the name...

Page 13: ...tilled water 2 Push the menu key and select SENSOR SETUP CALIBRATE 3 If the pass code is enabled in the security menu for the controller enter the pass code 4 Select ZERO CAL and push enter 5 Select t...

Page 14: ...rops of water Place the sensor vertically in the calibration cap with the membrane downwards Tighten the calibration cap and push enter 8 Wait for the value to stabilize and push enter 9 Review the ca...

Page 15: ...libration options on page 13 10 On the NEW SENSOR screen select whether the sensor is new Option Description YES The sensor was not calibrated previously with this controller The days of operation and...

Page 16: ...STANDBY Section 8 Maintenance W A R N I N G Multiple hazards Only qualified personnel must conduct the tasks described in this section of the document 8 1 Membrane replacement C A U T I O N Chemical e...

Page 17: ...embrane 1 Unscrew the filling screw and remove along with the gasket 2 Unscrew the holding nut 3 Carefully remove the electrode from the sensor body 4 Empty any electrolyte left in the sensor body 5 U...

Page 18: ...e and membrane visually check the silver anode If more than 2 3 of the surface is coated with silver bromide then an electrode rejuvenation is required To do this very softly polish the areas coated w...

Page 19: ...nd install finger tight The gold surface is scratched or damaged Change the electrode Sludge or particles on the cathode Clean the cathode with a soft and absorbent tissue Rinse the membrane Cable or...

Page 20: ...membrane filling screw and change the electrolyte and the membrane Electrolyte leaking Check the screw in parts membrane filling screw and change the electrolyte and the membrane Interferences mainly...

Page 21: ...number for the sensor module SENSOR INFORMATION Shows the name and serial number that was entered by the user CAL DAYS Shows the number of days since the last calibration CAL HISTORY Shows a list of t...

Page 22: ...he sensor CAL IN PROGRESS A calibration was started but not completed Return to calibration 9 3 2 Error list Errors may occur for various reasons An error icon consists of an exclamation point within...

Page 23: ...uaranty of the described product properties 1 IDENTIFICATION OF THE SUBSTANCE Product code 09181 A 3600 Product name Electrolyte 2 COMPOSITION INFORMATION ON INGREDIENTS Water 90 CAS No 108 18 9 Potas...

Page 24: ...cautions protective equipment and emergency procedures Use personal protective equipment Environmental precautions Do not flush into surface water or sanitary sewer system Methods and material for con...

Page 25: ...f Specific effects in experiment on an animal No data is available on the product itself 12 ECOLOGICAL INFORMATION Toxicity No data is available on the product itself Do not flush into surface water o...

Page 26: ...REGULATORY INFORMATION EU regulatory information The product does not need to be labelled in accordance with EC directives or respective national laws National regulatory information Water contaminati...

Page 27: ......

Page 28: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Reviews: