background image

User Manual

DOC022.52.80396

Pocket Pro

 Cond

LR

, Cond

HR

1

Summary of Contents for Pocket Pro CondHR

Page 1: ...User Manual DOC022 52 80396 Pocket Pro CondLR CondHR 1 ...

Page 2: ...onsequential damages resulting from any defect or omission in this manual The manufacturer reserves the right to make changes in this manual and the products it describes at any time without notice or obligation Revised editions are found on the manufacturer s website Safety information NOTIC E The manufacturer is not responsible for any damages due to misapplication or misuse of this product incl...

Page 3: ...plier for instructions on how to return end of life equipment producer supplied electrical accessories and all auxiliary items for proper disposal Certification Canadian Radio Interference Causing Equipment Regulation IECS 003 Class B Supporting test records reside with the manufacturer This Class B digital apparatus meets all requirements of the Canadian Interference Causing Equipment Regulations...

Page 4: ...ition the receiving antenna for the device receiving the interference 3 Try combinations of the above Product overview The Pocket Pro conductivityLR tester and conductivityHR tester measure the conductivity of general water samples Refer to Figure 1 These tester is waterproof and floats Figure 1 Product features 1 Sensor cap 3 Keypad 5 Lanyard attachment 2 Sensor 4 Display 6 Pocket clip Product co...

Page 5: ...emical type and are inserted in the correct orientation Do not mix new and used batteries WA RNING Fire hazard Battery substitution is not permitted Use only alkaline batteries Install the four AAA alkaline batteries in the tester Refer to the illustrated steps in Figure 3 Figure 3 Install the batteries User interface and navigation Display description Figure 4 shows the measurements calibration s...

Page 6: ...ry power is less than 10 Shows when the lock feature is on When the lock feature is on the parameter value on the display does not change Push to set the lock feature to on or off Shows when the tester is in calibration mode or when a calibration is due If shows next to the calibration icon the last calibration was not successful Keypad description Figure 5 shows the keypad Table 2 gives the key d...

Page 7: ...n standard CondLR tester One or two auto recognition calibration standards 1 Set the power to on 2 Remove the cap from the sensor 3 Push to go to calibration mode The auto recognition standard s the tester expects to measure shows on the bottom line 4 Rinse the sensor and cap with deionized water and blot dry 5 Pour the calibration standard shown into the cap to the fill line 6 Put the sensor full...

Page 8: ...mperature compensation factor linear 0 00 to 9 99 per C default 2 00 per C The correction factor may need to be identified experimentally For example the factor for ultrapure water is 4 55 per C and the factor for NaCl salt solution 2 125 per C AOFF Set the auto off feature to on default or off When set to on power is automatically set to off after 8 minutes of no activity rSEt Change the settings...

Page 9: ...rature range of the tester Contact technical support as necessary Note The tester can still be used if a temperature sensor failure has occurred but without automatic temperature compensation Battery icon flashes The batteries have less than 10 power remaining Replace all four batteries Refer to Install the batteries on page 5 Replacement parts and accessories WA RNING Personal injury hazard Use o...

Page 10: ...hach com www hach com HACH LANGE GMBH Willstätterstraße 11 D 40549 Düsseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE Sàrl 6 route de Compois 1222 Vésenaz SWITZERLAND Tel 41 22 594 6400 Fax 41 22 594 6499 Hach Company Hach Lange GmbH 2014 All rights reserved 12 2014 Edition 2 ...

Reviews: