73
Communication Menu
9.1.2 Example of use
In this example we use:
•
One PC with a RS232 port.
•
One "RS-485<->RS232 converter"
Procedure:
1.
Connect both RS-485 wires of the instrument to the "RS-485<->RS232 converter".
2.
Connect the "RS-485<->RS232 converter" to the PC RS232 port using a standard cable
(RS232 DB9 straight cable).
On the PC:
1.
Run "Hyperterminal" on the PC.
2.
Configure the PC COM port used (e.g. COM2). Menu "File/Properties/Configure".
3.
Configure the parameters "Baud rate", "Parity", "Nb of stop bits", "Byte size" (Menu
"File/Properties/Configure"). Use the same parameters for the instrument and the PC.
4.
Configure the "Font = Courier 10" (Menu "View/Font").
5.
Connect "Hyperterminal" (Menu "Call/Call").
6.
Save the data received in the file of your choice (Menu "Transfer/Capture Text/Start").
On the instrument:
1.
Use the menu "Communication/RS-485 Simple/Send files" and the button "All files".
When the transfer is finished, close the file with "Hyperterminal" (Menu "Transfer/Capture
Text/Stop"). Now, all the reports are saved in a text file on your PC.
Configuration report example
GENERAL CONFIGURATION
Measurement mode . . . . . . . . . Continuous mode
Pressure unit. . . . . . . . . . . . . . . bar
Temperature unit . . . . . . . . . . . . °C
Storage mode . . . . . . . . . . . . . . Rolling buffer
Storage RAM time. . . . . . . . . . . 10 [s]
Storage FLASH time . . . . . . . . . 3600 [s]
Autosave in FLASH. . . . . . . . . . Disabled
Channel 1
Sensor. . . . . . . . . . . . . . . . . . . . K1200-L
Medium . . . . . . . . . . . . . . . . . . . Liquid
Gas unit. . . . . . . . . . . . . . . . . . . ppm->ppb
Liquid . . . . . . . . . . . . . . . . . . . . Water
Resolution displayed. . . . . . . . . 1
Thermal alarm. . . . . . . . . . . . . . Enabled 90.0 [°C]
Measurement interval . . . . . . . . 5 [s]
Out of range protection . . . . . . . Enabled
User offset. . . . . . . . . . . . . . . . . Enabled
User offset value . . . . . . . . . . . . 45
Alarms
Low Low . . . . . . . . . . . . . . . . . . Enabled 100.000000 [ppb]
Low . . . . . . . . . . . . . . . . . . . . . . Enabled 200.000000 [ppb]
High. . . . . . . . . . . . . . . . . . . . . . Enabled 9000.000000 [ppb]
High High . . . . . . . . . . . . . . . . . Enabled 10000.000000 [ppb]
Hysteresis . . . . . . . . . . . . . . . . . 5 [%]
Delay. . . . . . . . . . . . . . . . . . . . . 15 [s]
Filter
State . . . . . . . . . . . . . . . . . . . . . Disabled
Type . . . . . . . . . . . . . . . . . . . . . Median
Depth . . . . . . . . . . . . . . . . . . . . 5
Central depth . . . . . . . . . . . . . . 0
Summary of Contents for ORBISPHERE K1200
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 39: ...38 Installation...
Page 45: ...44 User Interface...
Page 46: ...45 Section 5 View Menu Figure 34 View menu...
Page 49: ...48 View Menu...
Page 59: ...58 Calibration Menu...
Page 60: ...59 Section 8 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 62: ...61 Inputs Outputs Menu 8 4 Analog outputs Figure 41 Analog outputs menu...
Page 69: ...68 Inputs Outputs Menu...
Page 86: ...85 Section 11 Products Menu Figure 44 Products menu...
Page 89: ...88 Global Configuration Menu...
Page 90: ...89 Section 13 Services Menu Figure 46 Services menu Part 1...
Page 91: ...90 Services Menu Figure 47 Services menu Part 2...