1.
Go to the main menu, then select
Calibration
>
Matrix calibration
>
Start matrix calibration
.
Note: There can be one, two or three open matrix calibration points. One of the three open calibration points
must be deleted before a new calibration point can be added.
2.
Immediately follow the instructions on the controller to collect a grab sample. Accept the message
immediately after the grab sample is collected. Write the shown Point ID to the sample. Prepare
the grab sample (e.g., filtration) as soon as possible for laboratory measurement.
Note: The measurement data is initially saved on the controller for the probe, then adjusted when the
laboratory values are determined and entered.
3.
Measure the grab sample in the laboratory with the two parameters: NO
3
–N and NO
2
–N or NO
3
and NO
2
.
4.
Select
Calibration
>
Matrix calibration
>
Edit open calibration points
.
Note: The Edit open calibration points option only shows when the Matrix calibration is started.
Note: Make sure that the chemical formulas of the sensor and laboratory measurements are the same (e.g.,
NO
3
–N and NO
2
–N or NO
3
and NO
2
).
The first open calibration point is shown.
5.
If several grab samples are collected without the laboratory values entered in between, use the
Point ID to make sure that the assignment is correct.
a.
To enter an open calibration point, confirm the open point and then select
Enter lab values
if
the point ID matches the comparison measurement.
b.
To delete an open calibration point, confirm the open point and then select
Delete this point
.
6.
Select
Next
to move to the next point. Select
Back
to complete the matrix calibration.
7.
Select
Calibration
>
Matrix calibration
>
Information
.
Data about the actual matrix calibration shows.
8.
Select
Calibration
>
Matrix calibration
>
Calibration interval
to set a reminder for the
Matrix
calibration
.
Note: The calibration interval refers to the last completed matrix calibration.
If necessary, go back to the factory calibration. Select
Calibration
>
Matrix calibration
>
Reset to
default
.
5.4 Correction via Link2sc
The Link2sc procedure offers a secure method of data exchange between process probes and
Link2sc-compatible photometers using an SD memory card or via a local area network (LAN).
During a pure control measurement, the measurement data is transferred from the probe to the
photometer where it is then archived together with the photometric reference data that has been
recorded.
Refer to the Link2sc user manual for a detailed description of the Link2sc procedure.
5.5 Data logging
The SC Controller provides one data log and one event log for each instrument. The data log stores
the measurement data at selected intervals. The event log stores different types of events that occur
on the devices (i.e., configuration changes, alarms and warning conditions).
The data log and the event log can be saved. Refer to the SC Controller user manual for instructions.
Section 6 Maintenance
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
English
17
Summary of Contents for NT3100sc
Page 2: ...Table of Contents English 3 Fran ais 27 Espa ol 52 Portugu s 78 102 124 147 169 2...
Page 11: ...Figure 3 Bracket installation English 11...
Page 20: ...Figure 8 Replace wiper blade 1 mm and 2 mm Figure 9 Replace wiper blade 5 mm 20 English...
Page 86: ...Figura 3 Instala o do suporte 86 Portugu s...
Page 104: ...2 1 2 1 1 2 1 2 104...
Page 108: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5 mm 5 2 9 1 mm 2 mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 4 9 108...
Page 109: ...3 109...
Page 110: ...4 5 110...
Page 118: ...9 5 mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 118...
Page 126: ...2 1 1 2 1 2 126...
Page 130: ...2 6 2 2 1 NT3100sc NT3200sc 3 5 mm 5 9 2 9 1 mm 2 mm 5 4 LZY261 3 3 1 9 130...
Page 131: ...3 2 3 5 4 3 131...
Page 132: ...4 5 132...
Page 133: ...3 3 SC 6 6 4 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 133...
Page 140: ...8 1 mm 2 mm 9 5 mm 140...
Page 142: ...10 7 5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 UNSTABLE 142...
Page 143: ...5 6 NO3 NO3 N NO3 NO3 N NO3 NO3 N NO2 NO2 N NO2 NO2 N NO2 NO2 N 8 Web 143...
Page 149: ...2 1 1 2 1 2 UV 149...
Page 153: ...2 6 2 2 1 NT3100sc NT3200sc 3 9 5mm 5 2 9 1mm 2mm 5x 4 LZY261 3 3 1 OFF 3 2 3 5 9 153...
Page 154: ...4 3 154...
Page 155: ...4 5 155...
Page 156: ...3 3 SC 6 6 4 ON 5 5 1 5 2 1 SC200 2 06 2 SC200 SC1000 SC200 SC4500 Claros 3 156...
Page 163: ...9 5mm 6 4 MSDS SDS UV 1 2 3 4 5 6 7 a 10 b 163...
Page 164: ...c 10mL d 8 SC200 SC1000 SC4500 Claros Enter 9 NT3200sc NT3200sc NT3200sc 10 11 12 10 7 5 164...
Page 165: ...5 NO3 NO3 NO3 NO3 NO2 NO2 NO2 NO2 NOx NOx NOx NOx sc200 sc1000 6 165...
Page 171: ...5 1 2 PUR HT CE CMIM UKCA FCC ISED 1 2 2 2 1 2 1 1 171...
Page 172: ...2 1 2 UV 2 1 3 2 1 4 ICES 003 Class A Class A 172...
Page 175: ...1 1 3 2 2 6 2 2 1 NT3100sc NT3200sc 3 9 5 5 2 9 1 2 5 4 LZY261 9 175...
Page 176: ...3 3 1 OFF 3 2 3 5 4 176...
Page 177: ...3 177...
Page 178: ...4 5 178...
Page 186: ...9 5 6 4 MSDS SDS UV 1 Sensor setup 2 3 Maintenance 4 Hold 5 6 7 a 10 b 186...
Page 193: ......