Replace the conditioning solution
W A R N I N G
Inhalation hazard. Do not breathe Diisopropylamine (DIPA) or ammonia fumes. Exposure may result in
severe injury or death.
W A R N I N G
Chemical exposure hazard. Diisopropylamine (DIPA) and ammonia are a flammable, corrosive and
toxic chemical. Exposure may result in severe injury or death.
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Read the safety data sheet from the supplier
before bottles are filled or reagents are prepared. For laboratory use only. Make the hazard information
known in accordance with the local regulations of the user.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
Replace the conditioning solution at 3-month intervals or when less than 10% full. Use DIPA
(recommended) or ammonia (more than 28%) for the conditioning solution.
Items supplied by the user:
• Personal protective equipment (refer to MSDS/SDS)
• Diisopropylamine (DIPA) 99%, 1 L bottle
• Bottle adapter for Merck
®
or Orion
®
DIPA bottles if applicable
Replace the conditioning bottle as follows:
1.
Put on the personal protective equipment identified in the safety data sheet (MSDS/SDS).
2.
Push
menu
, then select STOP ANALYZER.
Note: If START ANALYZER shows, the analyzer is already in standby mode.
3.
Turn the latch on the analytic panel to the unlock position. Open the analytics panel.
4.
Replace the conditioning solution bottle. Refer to the illustrated steps that follow.
Note: Analyzers with the optional cationic pump have two tubes that go in the conditioning bottle cap. Remove
both tubes.
Note: At illustrated steps 1 and 8, only analyzers without an enclosure have a conditioning bottle holder.
Do illustrated steps 3–6 under a fume hood if available. Do not breathe DIPA or ammonia fumes.
6
English
Summary of Contents for NA5600 sc Na+
Page 7: ...English 7...
Page 13: ...Remove the reference electrode Refer to the illustrated steps that follow English 13...
Page 33: ...Deutsch 33...
Page 38: ...38 Deutsch...
Page 61: ...Italiano 61...
Page 66: ...66 Italiano...
Page 88: ...88 Fran ais...
Page 93: ...Fran ais 93...
Page 94: ...Retrait de l lectrode de r f rence Reportez vous aux tapes illustr es suivantes 94 Fran ais...
Page 115: ...Espa ol 115...
Page 120: ...120 Espa ol...
Page 142: ...142 Portugu s...
Page 147: ...Portugu s 147...
Page 148: ...Remover o el ctrodo de refer ncia Consulte os passos ilustrados abaixo 148 Portugu s...
Page 171: ...Nederlands 171...
Page 176: ...176 Nederlands...
Page 177: ...Referentie elektrode verwijderen Volg de volgende afgebeelde stappen Nederlands 177...
Page 198: ...198 Polski...
Page 203: ...Polski 203...
Page 204: ...Wyjmowanie elektrody odniesienia Sprawdzi list kolejnych czynno ci na rysunku 204 Polski...
Page 224: ...224 Svenska...
Page 230: ...Ta bort referenselektroden Se de illustrerade stegen som f ljer 230 Svenska...
Page 250: ...250 Suomi...
Page 256: ...Viite elektrodin irrottaminen Lue seuraavat kuvaohjeet 256 Suomi...
Page 277: ...Magyar 277...
Page 282: ...282 Magyar...
Page 283: ...A referenciaelektr da elt vol t sa L sd az al bbi k peken bemutatott l p seket Magyar 283...
Page 301: ...1 1 3 301 X 303 X KCl 306 X 307 X 1 2 MSDS SDS 3 10 301...
Page 303: ...1 DIPA DIPA 303...
Page 305: ...305...
Page 308: ...T 1 6 A 250 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 308...
Page 309: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 309...
Page 310: ...1 250 1 5 2x 2x KCl MSDS SDS 310...
Page 311: ...311...
Page 312: ...312...
Page 313: ...DIPA DIPA MSDS SDS 313...
Page 314: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 314...
Page 322: ...KCl2 1 4 KCl 322 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 2 KCl 1 322...
Page 323: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 323...
Page 335: ...T rk e 335...
Page 340: ...340 T rk e...
Page 341: ...Referans elektrotunun kar lmas A a da g sterilen resimli ad mlara bak n T rk e 341...
Page 359: ...1 1 3 359 X 361 X KCl 364 X 365 X 1 2 MSDS SDS 3 10 359...
Page 361: ...1 DIPA DIPA 361...
Page 363: ...363...
Page 366: ...T 1 6 A 250 VAC 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 366...
Page 367: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 367...
Page 368: ...1 250 mL 1 5 L 2x 2x KCl MSDS SDS 368...
Page 369: ...369...
Page 370: ...370...
Page 371: ...DIPA DIPA MSDS SDS 371...
Page 372: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 372...
Page 377: ...PH TOO LOW pH 10 361 o diag PERFORM TEST CONDITIONING VALVE 377...
Page 382: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 382...
Page 389: ......