Figure 6 Set the power switch to off
Discard the deionized water in the bottles
1.
Remove the reactivation solution and calibration solution bottles from the analyzer. Pull the bottle
caps down so they hang below the analyzer.
Do not let the bottle cap or internal tubing touch the analyzer to prevent contamination.
2.
Discard the deionized water in the bottles.
3.
Let the bottles fully air dry.
4.
Install the bottles in the analyzer.
5.
Close the analyzer door.
Troubleshooting
Problem
Possible cause
Solution
The status bar does
not show on the
measurement screen.
The channel shown on the
measurement screen is
not the channel being
measured.
Push the
UP
or
DOWN
arrow until the channel being
measured shows on the measurement screen.
A channel does not
show on the
measurement screen
when the
UP
or
DOWN
arrow key is pushed.
The channel is not
selected in the display
settings.
Push
menu
, then select SETUP SYSTEM > DISPLAY
SETUP > ADJUST ORDER > SEE CURRENT ORDER.
If the channel is not shown, push
back
, then select ADD
MEASUREMENTS. Push the
LEFT
arrow to select or
unselect a checkbox.
A channel is not being
measured.
The channel is not
activated.
Push
menu
, then select SETUP SYSTEM > CONFIGURE
ANALYZER > CONFIGURE SEQUENCER > ACTIVATE
CHANNELS. Add the channel. Push the
LEFT
arrow to
select or unselect a checkbox.
The channel is not in the
channel measurement
sequence.
Push
menu
, then select SETUP SYSTEM > CONFIGURE
ANALYZER > CONFIGURE SEQUENCER > SEQUENCE
CHANNELS.
Push the
UP
and
DOWN
arrows to select a row, then push
the
LEFT
or
RIGHT
arrow to select a channel. S1 is the
first channel measured followed by S2, S3 and S4.
Note: Do not select channels that contain the symbol "~" (e.g., 4-
~SAMPLE4). Channels that contain the symbol "~" are not
measured.
English
17
Summary of Contents for NA5600 sc Na+
Page 7: ...English 7...
Page 13: ...Remove the reference electrode Refer to the illustrated steps that follow English 13...
Page 33: ...Deutsch 33...
Page 38: ...38 Deutsch...
Page 61: ...Italiano 61...
Page 66: ...66 Italiano...
Page 88: ...88 Fran ais...
Page 93: ...Fran ais 93...
Page 94: ...Retrait de l lectrode de r f rence Reportez vous aux tapes illustr es suivantes 94 Fran ais...
Page 115: ...Espa ol 115...
Page 120: ...120 Espa ol...
Page 142: ...142 Portugu s...
Page 147: ...Portugu s 147...
Page 148: ...Remover o el ctrodo de refer ncia Consulte os passos ilustrados abaixo 148 Portugu s...
Page 171: ...Nederlands 171...
Page 176: ...176 Nederlands...
Page 177: ...Referentie elektrode verwijderen Volg de volgende afgebeelde stappen Nederlands 177...
Page 198: ...198 Polski...
Page 203: ...Polski 203...
Page 204: ...Wyjmowanie elektrody odniesienia Sprawdzi list kolejnych czynno ci na rysunku 204 Polski...
Page 224: ...224 Svenska...
Page 230: ...Ta bort referenselektroden Se de illustrerade stegen som f ljer 230 Svenska...
Page 250: ...250 Suomi...
Page 256: ...Viite elektrodin irrottaminen Lue seuraavat kuvaohjeet 256 Suomi...
Page 277: ...Magyar 277...
Page 282: ...282 Magyar...
Page 283: ...A referenciaelektr da elt vol t sa L sd az al bbi k peken bemutatott l p seket Magyar 283...
Page 301: ...1 1 3 301 X 303 X KCl 306 X 307 X 1 2 MSDS SDS 3 10 301...
Page 303: ...1 DIPA DIPA 303...
Page 305: ...305...
Page 308: ...T 1 6 A 250 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 308...
Page 309: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 309...
Page 310: ...1 250 1 5 2x 2x KCl MSDS SDS 310...
Page 311: ...311...
Page 312: ...312...
Page 313: ...DIPA DIPA MSDS SDS 313...
Page 314: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 314...
Page 322: ...KCl2 1 4 KCl 322 SET FUNCTION FEEDER CONTROL diag OVERFEED RESET 2 KCl 1 322...
Page 323: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 323...
Page 335: ...T rk e 335...
Page 340: ...340 T rk e...
Page 341: ...Referans elektrotunun kar lmas A a da g sterilen resimli ad mlara bak n T rk e 341...
Page 359: ...1 1 3 359 X 361 X KCl 364 X 365 X 1 2 MSDS SDS 3 10 359...
Page 361: ...1 DIPA DIPA 361...
Page 363: ...363...
Page 366: ...T 1 6 A 250 VAC 1 2 3 3 4 4 5 6 7 menu START ANALYZER 3 366...
Page 367: ...4 2 1 1 menu STOP ANALYZER START ANALYZER 2 KCl KCl KCl KCl 3 4 4 367...
Page 368: ...1 250 mL 1 5 L 2x 2x KCl MSDS SDS 368...
Page 369: ...369...
Page 370: ...370...
Page 371: ...DIPA DIPA MSDS SDS 371...
Page 372: ...1 MSDS SDS 2 5 3 4 5 6 menu REAGENTS STANDARDS PRIME REAGENTS 372...
Page 377: ...PH TOO LOW pH 10 361 o diag PERFORM TEST CONDITIONING VALVE 377...
Page 382: ...1 7 13 2 8 3 14 3 9 15 4 10 16 5 11 3 17 6 12 3 382...
Page 389: ......