![Hach MP-4 User Manual Download Page 33](http://html.mh-extra.com/html/hach/mp-4/mp-4_user-manual_577073033.webp)
Calibration
33
4.7
Calibrate resistivity
Resistivity is the reciprocal of conductivity. Resistivity is automatically calibrated
based on the solution type used during a conductivity calibration.
4.8
Reset factory calibration—conductivity,
mineral/salt or TDS
If calibration is suspect or known to be incorrect, and no standard solution is
available, it is possible to replace the calibrated value with the original factory value
for that solution. This ideal factory (
FAC
) value is the same for all MP Series
meters, and it returns to a known state without solution in the cup.
The
FAC
internal electronic calibration is not intended to replace calibration with
conductivity standard solutions.
1.
Push the
COND
key, the
MIN/SALT
key or the
TDS
key.
2.
Push the
CAL/MCLR
key twice in
COND
or three times in
TDS
.
Note:
In the User
mode, push the
CAL/MCLR
key twice in the
COND
mode and three
times in the
TDS
mode or the
MIN/SALT
mode. (This bypasses temperature correction
and ratio adjustments.)
3.
Push the
UP/MS
key until the
FAC
icon appears.
4.
Push the
CAL/MCLR
key
to accept the factory calibration setting.
5.
If another solution needs to be reset, select another solution type and repeat
the procedure.
4.9
pH calibration
Note:
Always zero-out the MP Series meter with a pH 7 buffer solution before calibrating with
acid or base buffers such as pH 4 or pH 10 solutions.
To perform a pH calibration:
1.
Rinse the sensor cups three times with a pH 7 buffer solution.
2.
Fill both sensor cups with pH 7 buffer solution.
3.
Push the
pH
key to verify the pH calibration. If the display shows 7.00, skip the
pH zero calibration and proceed to
Summary of Contents for MP-4
Page 2: ...2 ...
Page 6: ...Table of contents 6 ...
Page 10: ...Specifications 10 ...
Page 30: ...Operation 30 ...
Page 36: ...Calibration 36 ...
Page 42: ...Troubleshooting 42 ...
Page 48: ...Limited Warranty 48 ...
Page 52: ...Conductivity conversion 52 ...
Page 58: ...Additional information on pH and ORP MP 6 and MP 6p models 58 ...