Operation
18
2.
Hold the meter at a 30 degree angle and let the sample flow into the
conductivity cup continuously with no aeration.
3.
Push the
RES
key.
4.
Observe or record the value displayed.
Note:
If reading is lower than 10 k
Ω
, [- - - -] is shown. Measure conductivity for these samples.
3.7
Measure mineral/salt (MP-6p model only)
To measure mineral/salt:
1.
Rinse the conductivity cup three times with the sample to be measured. This
conditions the temperature compensation sensor and prepares the cell.
2.
Fill the conductivity cup with the solution.
3.
Push the
MIN/SALT
key.
4.
Observe or record the value displayed, or push
UP/MS
to store the reading.
3.8
Measure TDS
To measure TDS:
1.
Rinse the conductivity cup three times with the sample to be measured. This
conditions the temperature compensation sensor and prepares the cell.
2.
Fill the conductivity cup with the solution.
3.
Push the
TDS
key.
4.
Observe or record the value displayed, or push
UP/MS
to store the reading.
3.9
Measure ORP/Redox (MP-6 and MP-6p models)
To measure the ORP/Redox:
1.
Remove the protective pH/ORP sensor cap. Squeeze the sides and pull up.
2.
Rinse the sensor cup three times with the sample to be measured.
3.
Shake the meter after each rinse to remove residual liquid.
4.
Fill both sensor cups with the sample.
5.
Push the
ORP
key.
6.
Observe or record the value displayed, or push
UP/MS
to store the reading.
Important Note:
After the test, fill the pH/ORP sensor cup with pH Storage Solution
and replace the protective cap. Do not allow the pH/ORP sensor cup to dry out.
3.10 Measure pH (MP-6 and MP-6p models)
To measure pH:
1.
Remove the protective pH/ORP sensor cap. Squeeze the sides and pull up.
2.
Rinse the pH/ORP sensor cup three times with the sample to be measured.
3.
Shake the meter after each rinse to remove residual liquid.
4.
Fill both sensor cups with the sample.
Summary of Contents for MP-4
Page 2: ...2 ...
Page 6: ...Table of contents 6 ...
Page 10: ...Specifications 10 ...
Page 30: ...Operation 30 ...
Page 36: ...Calibration 36 ...
Page 42: ...Troubleshooting 42 ...
Page 48: ...Limited Warranty 48 ...
Page 52: ...Conductivity conversion 52 ...
Page 58: ...Additional information on pH and ORP MP 6 and MP 6p models 58 ...