4.
Select Parameter 1 and then Parameter 2 and select one of the
following options:
Option
Description
Parameter 1
mV, pH, ISE or Disabled
Parameter 2
EC, NaCl, TDS, Ω or Disabled
5.
Select Channel 2 and confirm with . Select % or mg/L and confirm.
Use a sample ID
The sample ID tag is used to associate readings with a particular sample
location. If assigned, stored data will include this ID.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Sample ID and confirm.
3.
Use the or to select
Option
Description
Automatic
A consecutive number will be automatically assigned to every
sample.
Manual
A keyboard or a barcode scanner is required to enter the
sample ID name before taking a measurement (maximum
15 characters).
Select the data output
Data can be stored or transferred to a printer or to a PC.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Data Output and confirm.
3.
Use the or to select
Option
Description
Deactivated
Select Deactivated if no printer or PC is connected.
Option
Description
For Printer
Select Dot matrix printer or Thermal printer.
For Computer
Select Terminal, LabCom or LabCom Easy. The LabCom
Software controls several modules, pH and conductivity
meters, automatic burettes, Samplers and so on from a
computer. The LabCom Easy software gets measurements
and calibration data from the meters
Note: Make sure that LabCom or LabCom Easy software has version 2.1 to be
compatible with the instrument.
Change the date and time
The date and time can be changed from the Date / Time menu.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Date / Time and confirm.
3.
Use the and to advance to the next step and use the or to
change a value. Confirm.
The current date and time will be shown on the display.
Adjust the display contrast
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Display contrast and confirm.
3.
Use the and to adjust the contrast of the display and confirm.
Adjust the temperature
The temperature measurement can be adjusted at 25 °C (77 °F) and/or
85 °C (185 °F) to increase accuracy.
1.
Put the probe and a reference thermometer in a container of water at
approximately 25 °C and allow the temperature to stabilize.
2.
Compare the temperature read by the meter with that of the
reference thermometer. The difference is the adjustment value for
the meter.
English
13
Summary of Contents for Lange SensION+ MM 378
Page 17: ...1 2 English 17 ...
Page 28: ...1 2 3 4 28 Deutsch ...
Page 40: ...1 2 40 Deutsch ...
Page 62: ...1 2 62 Italiano ...
Page 73: ...1 2 3 4 Français 73 ...
Page 84: ...1 2 84 Français ...
Page 95: ...1 2 3 4 Español 95 ...
Page 106: ...1 2 106 Español ...
Page 117: ...1 2 3 4 Português 117 ...
Page 128: ...1 2 128 Português ...
Page 150: ...1 2 150 Čeština ...
Page 161: ...1 2 3 4 Nederlands 161 ...
Page 173: ...1 2 Nederlands 173 ...
Page 194: ...1 2 194 Dansk ...
Page 205: ...1 2 3 4 Polski 205 ...
Page 216: ...1 2 216 Polski ...
Page 238: ...1 2 238 Svenska ...
Page 260: ...1 2 260 Suomi ...
Page 271: ...1 2 3 4 български 271 ...
Page 283: ...1 2 български 283 ...
Page 295: ...1 2 3 4 Magyar 295 ...
Page 307: ...1 2 Magyar 307 ...
Page 329: ...1 2 Română 329 ...
Page 351: ...1 2 lietuvių kalba 351 ...
Page 362: ...1 2 3 4 362 Русский ...
Page 374: ...1 2 374 Русский ...
Page 397: ...1 2 Türkçe 397 ...
Page 408: ...1 2 3 4 408 Slovenský jazyk ...
Page 419: ...1 2 Slovenský jazyk 419 ...
Page 441: ...1 2 Slovenski 441 ...
Page 463: ...1 2 Hrvatski 463 ...
Page 474: ...1 2 3 4 474 Ελληνικά ...
Page 487: ...1 2 Ελληνικά 487 ...
Page 511: ...1 2 eesti keel 511 ...
Page 518: ......
Page 519: ......