Hach H120 Operating Instructions Download Page 1

5600 Lindbergh Dr., Loveland, CO  80539
T

Te

ecch

hn

niicca

all  S

Su

up

pp

po

orrtt::

(800) 227-4224 (U.S.A. only)
Fax (970) 669-2932
[email protected] • www.hach.com

International customers:
Tel +001 (970) 669-3050
Fax +001 (970) 669-2932

[email protected]

Model H120

pH Meter with Silicon Chip Sensor

IQ Scientific Instruments

Calibration and Measurement Instructions

Parts

When finished, always rinse the probe and replace the protective cap.

C

Remove cap from probe
tip and turn on the meter.

Rinse the probe in tap
water and blot dry.

Apply one drop of 7.00
buffer to the probe tip.  Be
sure to cover both the
reference junction and the
pH sensor.

Depress the CAL 1 button
with the tip of the clip.

The CAL and 7.0 will flash
until a stable reading is
obtained.   When the CAL
disappears and 7.0 stops
blinking, calibration is
complete. 

Rinse the probe in tap
water and blot dry.
To measure the pH of a
sample, go to Step 7. 

Now measure the sample
pH by placing a single
drop on the probe tip or
dipping the probe tip
into solution.   

1

2

3

4

5

6

7

6. Protective Cap

7. Clip

8. Bubble

9. POWER Button

10. CAL 1 Button

1.

Probe Tip

pH sensor and reference electrode.

2.

pH Sensor

Fast response silicon chip pH sensor with built-in temperature sensor.

3.

Reference Electrode

Replaceable saturated KCl Ag/AgCl electrode.

4.

Reference Junction

Measurement is made when the reference junction and the pH sensor are
immersed in the solution.

5.

LCD Display

CAL 

(blinking): Calibration in process.  Measurement cannot be made.

CAL

(off):  Measurement can be made.

BAT

:  Low battery indicator.  Replace batteries.

6.

Protective Cap

Always replace the cap after use.

7.

Clip

Use the tip of the clip to depress the CAL button.

8.

Bubble

The reference solution bubble indicates the life of the reference electrode.

9.

POWER Button

Press to turn the meter on or off.

10. 

CAL 1 Button

Button for pH 7.0 calibration.

1. Probe Tip

2. pH Sensor

3. Reference Electrode

4. Reference Junction

5. LCD Display

OPERATING INSTRUCTIONS

pH Sensor

Reference Junction

DOC022.53.80059

mini_man_H120  9/8/09  9:26 AM  Page 1